    Fabrication of two-weir structure-based packed columns for on-chip solid-phase extraction of DNA

    No full text
    Microchip-based packed column SPE of DNA was performed using the microfabricated two-weir structure within a microchannel. We developed two methods to fabricate the two-weir structured glass chips: a "two-side etching/alignment" method and a simplified "one-side/two-step etching" method. The former rnethod required a straightforward alignment step, while the latter approach comprised a simplified wet etching process using paraffin wax as the temporary protective layer. Both methods were convenient and rapid as compared to the previous approaches. Through a reversibly sealed bead-introduction channel, beads can be fed into or out of the chip columns, thus enabling refreshment of the packing materials. Using the proposed chip columns, highly efficient,lambda-DNA extractions (average recovery > 80%) were performed with good chip-to-chip reproducibility (RSD < 10%). The on-chip SPE procedure was completed within 15 min at the flow rate of 3 mu L/min and the bulk of the loaded DNA was eluted within a small volume of similar to 8 mu L. Application of the microchip-based packed columns was demonstrated by purifying PCR-amplifiable genomic DNA from human hepatocellular carcinoma (HepG2) cells and human whole blood samples