Performance of the Two Aerogel Cherenkov Detectors of the JLab Hall A Hadron Spectrometer
We report on the design and commissioning of two silica aerogel Cherenkov
detectors with different refractive indices. In particular, extraordinary
performance in terms of the number of detected photoelectrons was achieved
through an appropriate choice of PMT type and reflector, along with some design
considerations. After four years of operation, the number of detected
photoelectrons was found to be noticeably reduced in both detectors as a result
of contamination, yellowing, of the aerogel material. Along with the details of
the set-up, we illustrate the characteristics of the detectors during different
time periods and the probable causes of the contamination. In particular we
show that the replacement of the contaminated aerogel and parts of the
reflecting material has almost restored the initial performance of the
detectors.Comment: 18 pages, 9 Figures, 4 Tables, 44 Reference