9 research outputs found

    Detection of Artificial Occlusal Caries in a Phosphor Imaging Plate System with Two Types of LCD Monitors Versus Three Different Films

    Full text link
    The aim of this study was to determine diagnostic performance of a storage phosphor plate system Digora® Optime (Soredex, Helsinki, Finland) with two types of LCD monitor in the detection of artificial caries when compared to Ultraspeed (D), Ektaspeed Plus (E), and Insight (F) radiographic films. Seventy extracted human molars—with artificial caries—were radiographed under identical standardized conditions using (1) a storage phosphor plate system Digora (Soredex, Helsinki, Finland), (2) Insight, (3) Ektaspeed Plus, and (4) Ultraspeed (Carestream Health Inc, Rochester, NY). All digital images and radiographs were examined by three observers for the presence or absence of artificial caries using a five-point confidence scale. Digital images were evaluated both on a LCD computer monitor (Philips 170S, Holland) and medical monitor—3 megapixel monochrome display (Me355i2, Totoku, Tokyo)—with brightness and contrast enhancement. Observer responses were evaluated using ROC analysis and other measurements for diagnostic accuracy. Storage phosphor images with medical monitor demonstrated higher mean Az values (0.70 ± 0.08) than digital images with computer monitor and conventional films. Storage phosphor images with medical monitor presented the highest score, 0.97, 0.90, 0.94, for each observer, respectively. Also, true positive observations (0.82) and positive likelihood ratios (2.71) were higher in enhanced storage phosphor images with medical monitor. Caries detection of mechanically created lesions by experienced radiologists is roughly comparable when examining D-speed film images and Digora images on both the computer and medical LCD monitors, and appears to be poorer on E- and F-speed film images
    corecore