21 research outputs found

    Simple noise reduction for DWI

    Get PDF
    Our purpose in this study was to reduce the noise in order to improve the SNR of Dw images with high b-value by using two correction schemes. This study was performed with use of phantoms made from water and sucrose at different concentrations, which were 10, 30, and 50 weight percent (wt%). In noise reduction for Dw imaging of the phantoms, we compared two correction schemes that are based on the Rician distribution and the Gaussian distribution. The highest error values for each concentration with use of the Rician distribution scheme were 7.3 % for 10 wt%, 2.4 % for 30 wt%, and 0.1 % for 50 wt%. The highest error values for each concentration with use of the Gaussian distribution scheme were 20.3 % for 10 wt%, 11.6 % for 30 wt%, and 3.4 % for 50 wt%. In Dw imaging, the noise reduction makes it possible to apply the correction scheme of Rician distribution

    Benefits of computed tomography in reducing mortality in emergency medicine

    Get PDF
    Performing accurate diagnosis using computed tomography (CT) in emergency medicine may reduce mortality rates in various diseases. In this observational, correlational and cross-sectional study, we conducted multiple regression analyses to investigate the relationship between CT utilization rates and mortality. In addition, we estimated the annual net profits from CT to show the profitability of introducing a CT system in each Japanese prefecture
    corecore