    Evaluation of two-dimensional electronic portal imaging device using integrated images during volumetric modulated arc therapy for prostate cancer

    Get PDF
    Background: The aim of the study was to evaluate analysis criteria for the identification of the presence of rectal gas during volumetric modulated arc therapy (VMAT) for prostate cancer patients by using electronic portal imaging device (EPID)-based in vivo dosimetry (IVD). Materials and methods: All measurements were performed by determining the cumulative EPID images in an integrated acquisition mode and analyzed using PerFRACTION commercial software. Systematic setup errors were simulated by moving the anthropomorphic phantom in each translational and rotational direction. The inhomogeneity regions were also simulated by the I’mRT phantom attached to the Quasar phantom. The presence of small and large air cavities (12 and 48 cm3) was controlled by moving the Quasar phantom in several timings during VMAT. Sixteen prostate cancer patients received EPID-based IVD during VMAT. Results: In the phantom study, no systematic setup error was detected in the range that can happen in clinical ( < 5-mm and < 3 degree). The pass rate of 2% dose difference (DD2%) in small and large air cavities was 98.74% and 79.05%, respectively, in the appearance of the air cavity after irradiation three quarter times. In the clinical study, some fractions caused a sharp decline in the DD2% pass rate. The proportion for DD2% < 90% was 13.4% of all fractions. Rectal gas was confirmed in 11.0% of fractions by acquiring kilo-voltage X-ray images after the treatment. Conclusions: Our results suggest that analysis criteria of 2% dose difference in EPID-based IVD was a suitable method for identification of rectal gas during VMAT for prostate cancer patients

    Setup accuracy and dose attenuation of a wooden immobilization system for lung stereotactic body radiotherapy

    Get PDF
    Background: We evaluated the setup error and dose absorption of an immobilization system with a shell and wooden baseplate (SW) for lung stereotactic body radiotherapy (SBRT). Materials and methods: Setup errors in 109 patients immobilized with an SW or BodyFix system (BF) were compared. Dose attenuation rates of materials for baseplates were measured with an ion-chamber. Ionization measurements were performed from 90° to 180° gantry angle in 10° increments, with the ball water equivalent phantom placed at the center of the wood and carbon baseplates whose effects on dose distribution were compared using an electron portal imaging device. Results: The ratio for the anterior-posterior, cranial-caudal, and right-left of the cases within 3-mm registered shifts in interfractional setup error were 90.9%, 89.2%, and 97.4% for the SW, and 93.2%, 91.6%, and 98.0% for the BF, respectively. For intrafractional setup error, 98.3%, 97.4%, and 99.1% for the SW and 96.6%, 95.8%, and 98.7% for the BF were within 3-mm registered shifts, respectively. In the center position, the average (minimum/maximum) dose attenuation rates from 90° to 180° for the wooden and carbon baseplates were 0.5 (0.1/2.8)% and 1.0 (–0.1/10.1)% with 6 MV, respectively. The gamma passing rates of 2%/2 mm for the wooden and carbon baseplates were 99.7% and 98.3% (p < 0.01). Conclusions: The immobilization system with an SW is effective for lung SBRT since it is comparable to the BF in setup accuracy. Moreover, the wooden baseplate had lower radiation attenuation rates and affected the dose distribution less than the carbon baseplate.

    Dosimetric impact of baseline drift in volumetric modulated arc therapy with breath holding

    No full text
    BackgroundWe investigated the change of dose distributions in volumetric modulated arc therapy (VMAT) under baseline drift (BD) during breath holding.Materials and methodsTen VMAT plans recalculated to a static field at a gantry angle of 0° were prepared for measurement with a 2D array device and five original VMAT plans were prepared for measurement with gafchromic films. These measurement approaches were driven by a waveform reproducing breath holding with BD. We considered breath holding times of 15 and 10 s, and BD at four speeds; specifically, BD0 (0 mm/s), BD0.2 (0.2 mm/s), BD0.3 (0.3 mm/s), and BD0.4 (0.4 mm/s). The BD was periodically reproduced from the isocenter along the craniocaudal direction and the shift during breath holding (ShiftBH) ranged 0–6 mm.The dose distribution of BD0.2, BD0.3 and BD0.4 were compared to that of BD0 using gamma analysis with the criterion of 2%/2 mm.ResultsThe mean pass rates of each ShiftBH were 99.8% and 98.9% at 0 mm, 96.8% and 99.4% at 2 mm, 94.9% and 98.6% at 3 mm, 91.5% and 98.4% at 4 mm, 70.8% and 94.1% at 4.5 mm, and 55.0% and 83.6% at 6 mm for the array and film measurements, respectively.ConclusionWe found significant differences in ShiftBH above 4 mm (ρ

    Comparison of interfractional setup reproducibility between two types of patient immobilization devices in image-guided radiation therapy for prostate cancer

    No full text
    Purpose: The aim of this study is to compare the interfractional setup reproducibility of two types of patient immobilization devices for prostate cancer receiving image-guided radiation therapy (IGRT). Materials and Methods: The MOLDCARE (MC) involves hydraulic fixation, whereas the BlueBAG (BB) and Vac-Lock (VL) involve vacuum fixation. For 72 patients, each immobilization device was individually customized during computed tomography (CT) simulation. Before the treatment, bony registration was performed using orthogonal kV images and digitally reconstructed radiographs. The shift of the treatment couch was recorded as a benchmark in the first session. In subsequent sessions, the shifts from the benchmark were measured and analyzed. Soft-tissue registration was performed weekly by cone-beam CT and CT images, and the shifts were measured and analyzed. Results: In the superior-inferior and left-right directions, there were nearly no changes in the overall mean among the immobilization devices. In the anterior-posterior (AP) direction, the overall mean for the MC, BB, and VL were 0.34 ± 1.33, −0.47 ± 1.27, and −1.82 ± 1.65 mm, respectively. The mean shifts along the AP direction were approximately 1 mm more in patients immobilized on the BB and 2.5 mm more in those on the VL, compared to those on the MC, after the twentieth treatment. No significant changes were observed among the patients immobilized on those devices, respectively, in soft-tissue registration. Conclusion: It can be concluded that the settling of the vacuum fixation was caused by air leakage in the latter-half treatment, and the immobilization device type has no effect on the treatment-position reproducibility in IGRT

    Characterization of knowledge-based volumetric modulated arc therapy plans created by three different institutions’ models for prostate cancer

    No full text
    BackgroundThe aim of this study was to clarify factors predicting the performance of knowledge-based planning (KBP) models in volume modulated arc therapy for prostate cancer in terms of sparing the organ at risk (OAR).Materials and methodsIn three institutions, each KBP model was trained by more than 20 library plans (LP) per model. To validate the characterization of each KBP model, 45 validation plans (VP) were calculated by the KBP system. The ratios of overlap between the OAR volume and the planning target volume (PTV) to the whole organ volume (Voverlap/Vwhole) were analyzed for each LP and VP. Regression lines between dose–volume parameters (V90, V75, and V50) and Voverlap/Vwhole were evaluated. The mean OAR dose, V90, V75, and V50 of LP did not necessarily match those of VP.ResultsIn both the rectum and bladder, the dose–volume parameters for VP were strongly correlated with Voverlap/Vwhole at institutes A, B, and C (R > 0.74, 0.85, and 0.56, respectively). Except in the rectum at institute B, the slopes of the regression lines for LP corresponded to those for VP. For dose–volume parameters for the rectum, the ratios of slopes of the regression lines in VP to those in LP ranged 0.51–1.26. In the bladder, most ratios were less than 1.0 (mean: 0.77).ConclusionFor each OAR, each model made distinct dosimetric characterizations in terms of Voverlap/Vwhole. The relationship between dose–volume parameters and Voverlap/Vwhole of OARs in LP predicts the KBP models’ performance sparing OARs