New design of electrostatic mirror actuators for application in high-precision interferometry
We describe a new geometry for electrostatic actuators to be used in sensitive laser interferometers, suited for prototype and table top experiments related to gravitational wave detection with mirrors of 100 g or less. The arrangement consists of two plates at the sides of the mirror (test mass), and therefore does not reduce its clear aperture as a conventional electrostatic drive (ESD) would do. Using the sample case of the AEI-10 m prototype interferometer, we investigate the actuation range and the influence of the relative misalignment of the ESD plates with respect to the test mass. We find that in the case of the AEI-10 m prototype interferometer, this new kind of ESD could provide a range of 0.28 μm when operated at a voltage of 1 kV. In addition, the geometry presented is shown to provide a reduction factor of about 100 in the magnitude of the actuator motion coupling to the test mass displacement. We show that therefore in the specific case of the AEI-10 m interferometer, it is possible to mount the ESD actuators directly on the optical table without spoiling the seismic isolation performance of the triple stage suspension of the main test masses
Optimal time-domain combination of the two calibrated output quadratures of GEO 600
GEO 600 is an interferometric gravitational wave detector with a 600 m arm-length and which uses a dual-recycled optical configuration to give enhanced sensitivity over certain frequencies in the detection band. Due to the dual-recycling, GEO 600 has two main output signals, both of which potentially contain gravitational wave signals. These two outputs are calibrated to strain using a time-domain method. In order to simplify the analysis of the GEO 600 data set, it is desirable to combine these two calibrated outputs to form a single strain signal that has optimal signal-to-noise ratio across the detection band. This paper describes a time-domain method for doing this combination. The method presented is similar to one developed for optimally combining the outputs of two colocated gravitational wave detectors. In the scheme presented in this paper, some simplifications are made to allow its implementation using time-domain methods
Reducing Thermal Noise in Future Gravitational Wave Detectors by employing Khalili Etalons
Reduction of thermal noise in dielectric mirror coatings is a key issue for
the sensitivity improvement in second and third generation interferometric
gravitational wave detectors. Replacing an end mirror of the interferometer by
an anti-resonant cavity (a so-called Khalili cavity) has been proposed to
realize the reduction of the overall thermal noise level. In this article we
show that the use of a Khalili etalon, which requires less hardware than a
Khalili cavity, yields still a significant reduction of thermal noise. We
identify the optimum distribution of coating layers on the front and rear
surfaces of the etalon and compare the total noise budget with a conventional
mirror. In addition we briefly discuss advantages and disadvantages of the
Khalili etalon compared with the Khalili cavity in terms of technical aspects,
such as interferometric length control and thermal lensing.Comment: 13 pages, 9 figure
Triple Michelson Interferometer for a Third-Generation Gravitational Wave Detector
The upcoming European design study `Einstein gravitational-wave Telescope'
represents the first step towards a substantial, international effort for the
design of a third-generation interferometric gravitational wave detector. It is
generally believed that third-generation instruments might not be installed
into existing infrastructures but will provoke a new search for optimal
detector sites. Consequently, the detector design could be subject to fewer
constraints than the on-going design of the second generation instruments. In
particular, it will be prudent to investigate alternatives to the traditional
L-shaped Michelson interferometer. In this article, we review an old proposal
to use three Michelson interferometers in a triangular configuration. We use
this example of a triple Michelson interferometer to clarify the terminology
and will put this idea into the context of more recent research on
interferometer technologies. Furthermore the benefits of a triangular detector
will be used to motivate this design as a good starting point for a more
detailed research effort towards a third-generation gravitational wave
detector.Comment: Minor corrections to the main text and two additional appendices. 14
pages, 6 figure
Microelectromechanical system gravimeters as a new tool for gravity imaging
A microelectromechanical system (MEMS) gravimeter has been manufactured with a sensitivity of 40 ppb in an integration time of 1 s. This sensor has been used to measure the Earth tides: the elastic deformation of the globe due to tidal forces. No such measurement has been demonstrated before now with a MEMS gravimeter. Since this measurement, the gravimeter has been miniaturized and tested in the field. Measurements of the free-air and Bouguer effects have been demonstrated by monitoring the change in gravitational acceleration measured while going up and down a lift shaft of 20.7 m, and up and down a local hill of 275 m. These tests demonstrate that the device has the potential to be a useful field-portable instrument. The development of an even smaller device is underway, with a total package size similar to that of a smartphone
The upgrade of GEO600
The German / British gravitational wave detector GEO 600 is in the process of
being upgraded. The upgrading process of GEO 600, called GEO-HF, will
concentrate on the improvement of the sensitivity for high frequency signals
and the demonstration of advanced technologies. In the years 2009 to 2011 the
detector will undergo a series of upgrade steps, which are described in this
paper.Comment: 9 pages, Amaldi 8 conference contributio
- …