    Neointimal characteristics comparison between biodegradable-polymer and durable-polymer drug-eluting stents: 3-month follow-up optical coherence tomography light property analysis from the RESTORE registry

    Get PDF
    We aimed to quantitatively assess a possible difference of the neointimal quality between biodegradable polymer- (BP-) and durable polymer drug-eluting stents (DP-DESs). We conducted a single-center all-comer prospective cohort study: the RESTORE registry (UMIN000033009). All patients who received successful OCT examination at planned 3-month follow-up after DES implantation were analyzed. Study population was divided into 2 groups, BP-DES versus DP-DES groups. We evaluated standard OCT variables, coverage percent, and the quantitative light property values including light intensity, attenuation, and backscatter. We performed OCT analyses of 121 lesions in 98 patients (BP-DES 55 lesions in 51 patients vs. DP-DES 66 lesions in DP-DES 53 patients). Lesion and procedural characteristics were overall well-balanced between both groups. At 3-month follow-up, neointimal thickness (BP-DES 49.3 [38.2, 57.7] µm versus DP-DES 54.7 [45.1, 70.7] µm, p = 0.059) and coverage percent (BP-DES 94.5 [89.8, 97.0]% vs. DP-DES 95.8 [91.1, 98.1]%, p = 0.083) did not significantly differ. Light intensity of superficial neointima in the BP-DES was lower than that in the DP-DES, whereas that of deep neointima did not differ between both groups