'Office of Scientific and Technical Information (OSTI)'
Publication date
31/10/1971
Field of study
A gas chromatograph has been assembled which provides computer control of sample injection, column temperature, and flow rate, plus direct computer readout of inlet pressure, mass flow rate, and detector response. Data processing yields, in real-time, a standard deviation of less than 0.05% in retention time, which is comparable to previous results obtained using an off-line computer. However, corrected retention volumes determined in real-time had a standard deviation of about 0.4% which reflected primarily the uncertainty in flow measurement