Dosimetry characterization of P intravascular brachytherapy source wires using Monte Carlo codes PENELOPE and GEANT4
Monte Carlo calculations using the codes PENELOPE and GEANT4 have been
performed to characterize the dosimetric parameters of the new 20 mm long
catheter based P beta source manufactured by Guidant Corporation. The
dose distribution along the transverse axis and the two dimensional dose rate
table have been calculated. Also, the dose rate at the reference point, the
radial dose function and the anisotropy function were evaluated according to
the adapted TG-60 formalism for cylindrical sources. PENELOPE and GEANT4 codes
were first verified against previous results corresponding to the old 27 mm
Guidant P beta source. The dose rate at the reference point for the
unsheathed 27 mm source in water was calculated to be cGy
s mCi, for PENELOPE, and cGy s
mCi, for GEANT4. For the unsheathed 20 mm source these values were
cGy s mCi and cGy s
mCi, respectively. Also, a comparison with the limited data available on
this new source is shown. We found non negligible differences between the
results obtained with PENELOPE and GEANT4.Comment: 13 pages, 7 figures, 7 tables (accepted for publication in Medical