    Contribution à l'étude du phénomène de vieillissement des compteurs Geiger-Muller autocoupants

    No full text
    As an unambigous criterion for determining the deterioration of a G. M. couter tube, the value ΔVI/ΔVH, the change in pulse-hight as a function of the applied high-voltage, has been defined. A study was made of the relative importance of the three main constituents of a G. M. counter on its characteristics. Asystematic study was made of the influence of anode, cathode and of the gas-phase composition. It is shown that the bad functioning of an aged counter can be ascribed to two different factors : changes of the gas phase, that influence mainly the plateau characteristics and changes on the anode surface that cause a gradual decrease of the value Δ VI/ΔVH. This is taken as evidence that the changes of the anode surface influence the multiplication process.La valeur ΔVI/ΔVH, le changement de la hauteur des impulsions en fonction de la haute tension appliquée, a été déterminée comme critère sans équivoque pour l'appréciation du comportement d'un compteur de Geiger autocoupant. L'influence de l'état des trois facteurs essentiels, la phase gazeuse, l'anode et la cathode sur le fonctionnernent d'un tube a été étudiée. On a pu démontrer que la mise hors d'usage d'un compteur est due aux facteurs suivants : changements dans la phase gazeuse qui influencent principalement les caractéristiques du plateau et le changement dans l'état de l'anode qui influence surtout la valeur de Δ VI/ΔVH, c'est-à-dire le procès de multiplication