1 research outputs found

    Evaluation of the highly sensitive Roche thyroglobulin II assay and establishment of a reference limit for thyroglobulin-negative patient samples

    Get PDF
    Objectives: Thyroglobulin (Tg) measurements are used to monitor for residual thyroid tissue in patients with differentiated thyroid cancer (DTC) after thyroidectomy and radioiodine ablative therapy. In recent years highly sensitive Tg assays have been developed. In this study the analytical performance of the new Roche Elecsys Tg II assay was evaluated and compared with the well documented Access2 Tg assay (Beckman–Coulter). Design and methods: Analytical performance was examined using various Clinical and Laboratory Standards Institute (CLSI) evaluation protocols. Tg negative patient sera were used to establish an upper reference limit (URL) for the Elecsys Tg II assay. Results: Non-linearity, drift and carry-over according to CLSI EP10 and EP6 in a measuring range of 0.04–500 ng/mL were non-significant. Total precision according to CLSI EP5 was 10% at a Tg concentration of 0.08 ng/mL. A patient serum comparison performed according to a modified CLSI EP9 protocol showed a significant difference of a factor of approximately 1.4, despite using an identical CRM calibrator. The Elecsys Tg II assay measured Tg with a two-fold higher sensitivity than the Access2 assay. Finally, using human sera without Tg, an URL of 0.05 ng/mL was determined. Conclusions: In our hands the highly sensitive Elecsys Tg II assay shows a good analytical performance and a higher sensitivity compared to the Access2 Tg assay. An URL of 0.05 ng/mL for the Elecsys Tg II assay was determined which may improve the clinical utility of the assay for the detection of residual DTC or disease recurrence. Keywords: Thyroglobulin, Roche Elecsys Tg II assay, validation, reporting limi
    corecore