3 research outputs found
Antibacterial Envelope to Prevent Cardiac Implantable Device Infection
BACKGROUND
Infections after placement of cardiac implantable electronic devices (CIEDs) are associated with substantial morbidity and mortality. There is limited evidence on prophylactic strategies, other than the use of preoperative antibiotics, to prevent such infections.
METHODS
We conducted a randomized, controlled clinical trial to assess the safety and efficacy of an absorbable, antibiotic-eluting envelope in reducing the incidence of infection associated with CIED implantations. Patients who were undergoing a CIED pocket revision, generator replacement, or system upgrade or an initial implantation of a cardiac resynchronization therapy defibrillator were randomly assigned, in a 1:1 ratio, to receive the envelope or not. Standard-of-care strategies to prevent infection were used in all patients. The primary end point was infection resulting in system extraction or revision, long-term antibiotic therapy with infection recurrence, or death, within 12 months after the CIED implantation procedure. The secondary end point for safety was procedure-related or system-related complications within 12 months.
RESULTS
A total of 6983 patients underwent randomization: 3495 to the envelope group and 3488 to the control group. The primary end point occurred in 25 patients in the envelope group and 42 patients in the control group (12-month Kaplan-Meier estimated event rate, 0.7% and 1.2%, respectively; hazard ratio, 0.60; 95% confidence interval [CI], 0.36 to 0.98; P = 0.04). The safety end point occurred in 201 patients in the envelope group and 236 patients in the control group (12-month Kaplan-Meier estimated event rate, 6.0% and 6.9%, respectively; hazard ratio, 0.87; 95% CI, 0.72 to 1.06; P<0.001 for noninferiority). The mean (±SD) duration of follow-up was 20.7±8.5 months. Major CIED-related infections through the entire follow-up period occurred in 32 patients in the envelope group and 51 patients in the control group (hazard ratio, 0.63; 95% CI, 0.40 to 0.98).
CONCLUSIONS
Adjunctive use of an antibacterial envelope resulted in a significantly lower incidence of major CIED infections than standard-of-care infection-prevention strategies alone, without a higher incidence of complications. (Funded by Medtronic; WRAP-IT ClinicalTrials.gov number, NCT02277990.)
Safety, efficacy, and reliability evaluation ofa novel small-diameter defibrillation lead: Global LEADR pivotal trial results.
BACKGROUND: Implantable cardioverter-defibrillators last longer, and interest in reliable leads with targeted lead placement is growing. The OmniaSecure defibrillation lead is a novel, small-diameter, catheter-delivered lead designed for targeted placement, based on the established SelectSecure SureScan MRI Model 3830 lumenless pacing lead platform.
OBJECTIVE: This trial assessed safety and efficacy of the OmniaSecure defibrillation lead.
METHODS: The worldwide LEADR pivotal clinical trial enrolled patients indicated for de novo implantation of a primary or secondary prevention implantable cardioverter-defibrillator or cardiac resynchronization therapy defibrillator, all of whom received the study lead. The primary efficacy end point was successful defibrillation at implantation per protocol. The primary safety end point was freedom from study lead-related major complications at 6 months. The primary efficacy and safety objectives were met if the lower bound of the 2-sided 95% credible interval was \u3e88% and \u3e90%, respectively.
RESULTS: In total, 643 patients successfully received the study lead, and 505 patients have completed 12-month follow-up. The lead was placed in the desired right ventricular location in 99.5% of patients. Defibrillation testing at implantation was completed in 119 patients, with success in 97.5%. The Kaplan-Meier estimated freedom from study lead-related major complications was 97.1% at 6 and 12 months. The trial exceeded the primary efficacy and safety objective thresholds. There were zero study lead fractures and electrical performance was stable throughout the mean follow-up of 12.7 ± 4.8 months (mean ± SD).
CONCLUSION: The OmniaSecure lead exceeded prespecified primary end point performance goals for safety and efficacy, demonstrating high defibrillation success and a low occurrence of lead-related major complications with zero lead fractures
Antibacterial Envelope to Prevent Cardiac Implantable Device Infection
Background Infections after placement of cardiac implantable electronic devices (CIEDs) are associated with substantial morbidity and mortality. There is limited evidence on prophylactic strategies, other than the use of preoperative antibiotics, to prevent such infections. Methods We conducted a randomized, controlled clinical trial to assess the safety and efficacy of an absorbable, antibiotic-eluting envelope in reducing the incidence of infection associated with CIED implantations. Patients who were undergoing a CIED pocket revision, generator replacement, or system upgrade or an initial implantation of a cardiac resynchronization therapy defibrillator were randomly assigned, in a 1:1 ratio, to receive the envelope or not. Standard-of-care strategies to prevent infection were used in all patients. The primary end point was infection resulting in system extraction or revision, long-term antibiotic therapy with infection recurrence, or death, within 12 months after the CIED implantation procedure. The secondary end point for safety was procedure-related or system-related complications within 12 months. Results A total of 6983 patients underwent randomization: 3495 to the envelope group and 3488 to the control group. The primary end point occurred in 25 patients in the envelope group and 42 patients in the control group (12-month Kaplan-Meier estimated event rate, 0.7% and 1.2%, respectively; hazard ratio, 0.60; 95% confidence interval [CI], 0.36 to 0.98; P=0.04). The safety end point occurred in 201 patients in the envelope group and 236 patients in the control group (12-month Kaplan-Meier estimated event rate, 6.0% and 6.9%, respectively; hazard ratio, 0.87; 95% CI, 0.72 to 1.06; P<0.001 for noninferiority). The mean (+/- SD) duration of follow-up was 20.7 +/- 8.5 months. Major CIED-related infections through the entire follow-up period occurred in 32 patients in the envelope group and 51 patients in the control group (hazard ratio, 0.63; 95% CI, 0.40 to 0.98). Conclusions Adjunctive use of an antibacterial envelope resulted in a significantly lower incidence of major CIED infections than standard-of-care infection-prevention strategies alone, without a higher incidence of complications