Evaluation of a new rebound tonometer for self-measurement of intraocular pressure
BACKGROUND/AIMS: To compare the accuracy of self-obtained, partner-obtained and trainer-obtained measurements using the handheld Icare Home rebound tonometer with Goldmann applanation tonometry (GAT), and to evaluate the acceptability to subjects of Icare Home measurement.
METHODS: 76 subjects were trained to use Icare Home for self-measurement using a standardised protocol. A prespecified checklist was used to assess the ability of a subject to perform self-tonometry. Accuracy of Icare Home self-measurement was compared with GAT using one eye per subject, randomly selected. Bland-Altman difference analysis was used to compare Icare Home and GAT intraocular pressure (IOP) estimates. Acceptability of self-tonometry was evaluated using a questionnaire.
RESULTS: 56 subjects (74%, 95% CI 64 to 84) were able to correctly perform self-tonometry. Mean bias (95% limits of agreement) was 0.3â
mmâ
Hg (â4.6 to 5.2), 1.1â
mmâ
Hg (â3.2 to 5.3) and 1.2â
mmâ
Hg (â3.9 to 6.3) for self-assessment, partner-assessment and trainer-assessment, respectively, suggesting underestimation of IOP by Icare Home tonometry. Differences between GAT and Icare Home IOP were greater for central corneal thickness below 500â
”m and above 600â
”m than data points within this range. Acceptability questionnaire responses showed high agreement that the self-pressure device was easy to use (84%), the reading was quick to obtain (88%) and the measurement was comfortable (95%).
CONCLUSIONS: Icare Home tonometry can be used for self-measurement by a majority of trained subjects. IOP measurements obtained using Icare Home tonometry by self-assessment and third party-assessment showed slight underestimation compared with GAT
Recommended from our members
Impact of optical coherence tomography on diagnostic decision-making by UK community optometrists: a clinical vignette study.
PURPOSE: In recent years, there has been widespread investment in imaging technologies by community optometrists in the UK, most notably optical coherence tomography (OCT). The aim of the current study was to determine the value of OCT in the diagnosis of posterior segment diseases in a representative sample of community optometrists using a clinical vignette methodology.
METHODS: A group of community optometrists (n = 50) initially completed a standardised training package on OCT interpretation followed by a computer-based assessment featuring 52 clinical vignettes, containing images of healthy (n = 8) or glaucomatous (n = 18) discs or healthy (n = 8) or diseased (n = 18) fundi. Each vignette featured either a single fundus/disc photographic image, or a combination of a fundus/disc image with the corresponding OCT scan. An expert panel confirmed that the fundus images presented alone and those in combination with OCT data were of a similar level of difficulty and that the cases were typical of those seen in primary care. For each case, the optometrist selected their diagnosis from a pull-down list and reported their confidence in their decision using a 10-point Likert scale. Pairwise comparisons of the fundus image alone and fundus image/OCT combination were made for both diagnostic performance and confidence.
RESULTS: The mean percentage of correct diagnoses using fundus imaging alone was 62% (95% CI 59-64%) and for the combination of fundus image/OCT was 80% (95% CI 77-82%). The mean false negative rate with fundus alone was 27% reducing to 13% with the OCT combination. Median confidence scores for fundus imaging alone was 8.0 (IQR 7.0-8.0) and 8.3 (IQR 8.0-9.0) for the combination. Improvements in performance and confidence were statistically significant (p < 0.001).
CONCLUSION: The results from this vignette study suggests that OCT improves optometrists' diagnostic performance compared to fundus observation alone. These initial results suggest that OCT provides valuable additional data that could augment case-finding for glaucoma and retinal disease; however, further research is needed to assess its diagnostic performance in a routine clinical practice setting
Simplified automatic method for measuring the visual field using the perimeter ZERK 1
Background: Currently available perimeters have limited capabilities of performing measurements of the visual field in children. In addition, they do not allow for fully automatic measurement even in adults. The patient in each case (in any type of perimeter) has at his disposal a button which he uses to indicate that he has seen a light stimulus. Such restrictions have been offset in the presented new perimeter ZERK 1. Methods: The paper describes a new type of automated, computerized perimeter designed to test the visual field in children and adults. The new perimeter and proprietary software enable to carry out tests automatically (without the need to press any button). The presented full version of the perimeter has been tested on a head phantom. The next steps will involve clinical trials and a comparison with measurements obtained using other types of perimeters. Results: The perimeter ZERK 1 enables automatic measurement of the visual field in two axes (with a span of 870 mm and a depth of 525 mm) with an accuracy of not less than 1o (95 LEDs on each arm) at a typical position of the patient's head. The measurement can be carried out in two modes: default/typical (lasting about 1 min), and accurate (lasting about 10 min). Compared with available and known types of perimeters, it has an open canopy, proprietary software and cameras tracking the eye movement, automatic control of fixation points, light stimuli with automatically preset light stimulus intensity in the following ranges: 550-700 mcd (red 620-630 nm), 1100-1400 mcd (green 515-530 nm), 200-400 mcd (blue 465-475 nm). Conclusions: The paper presents a new approach to the construction of perimeters based on automatic tracking of the eye movements in response to stimuli. The unique construction of the perimeter and the software allow for its mobile use in the examination of children and bedridden patients