Comparison Between the CASIA SS-1000 and Pentacam in Measuring Corneal Curvatures and Corneal Thickness Maps
PURPOSE: To compare the intra-device repeatability and inter-device reproducibility between two anterior segment imaging instruments, the CASIA SS-1000 (Tomey Corp., Nagoya, Japan) and Pentacam (OCULUS, Arlington, WA) in measuring anterior segment parameters.
METHODS: Single-center, prospective clinical trial. Participants ≥20 years of age were included. One eye was randomly selected, each imaged by three CASIA SS-1000 devices and three Pentacam devices by three different examiners. Each photographer operated a pair of devices, one CASIA SS-1000 and one Pentacam. The image order for each participant was determined by a random permutation table. Three images were taken from each device. A total of 18 images were taken for each eye. Ratios of the standard deviations, referenced as (CASIA/Pentacam), were calculated to compare the repeatability and reproducibility of the two imaging instruments.
RESULTS: In all, 66 participants with a mean age of 46.4 years (±21.7) were enrolled in the study. All repeatability ratios and intra-device variability were less than 1 (anterior corneal curvature: flat = 0.86, steep = 0.85; posterior corneal curvature: flat = 0.43, steep = 0.61; and map: thinnest = 0.22; central = 0.24, 2 mm = 0.26, 4 mm = 0.27, and 6 mm = 0.30). All reproducibility ratios, which measure the inter-device variability, were less than 1 (anterior corneal curvature: flat = 0.58, steep = 0.73; posterior corneal curvature: flat = 0.25, steep = 0.31; and pachymetry map: thinnest = 0.20; central = 0.20; 2 mm = 0.20; 4 mm = 0.19; and 6 mm = 0.22). A ratio of less than 1 indicates that the CASIA SS-1000 has more consistent measurements.
CONCLUSIONS: The CASIA SS-1000 was found to have better repeatability and reproducibility compared to the Pentacam for both corneal curvature and pachymetry maps. This greater consistency may require further study to determine whether the decreased variability can be translated into improved clinical results
Sources of Calcium at Connexin 36 Gap Junctions in the Retina
Synaptic plasticity is a fundamental feature of the CNS that controls the magnitude of signal transmission between communicating cells. Many electrical synapses exhibit substantial plasticity that modulates the degree of coupling within groups of neurons, alters the fidelity of signal transmission, or even reconfigures functional circuits. In several known examples, such plasticity depends on calcium and is associated with neuronal activity. Calcium-driven signaling is known to promote potentiation of electrical synapses in fish Mauthner cells, mammalian retinal AII amacrine cells, and inferior olive neurons, and to promote depression in thalamic reticular neurons. To measure local calcium dynamic
Postoperative Complications of True Dropless Cataract Surgery versus Standard Topical Drops
Purpose  Compare postoperative outcomes in cataract surgery between eyes with standard drop regimen versus dropless protocol by residents.
Design  Retrospective cohort study between April 1, 2018 and March 31, 2020.
Methods  The study was performed at Lyndon B. Johnson General Hospital in Houston, Harris County, Texas. A total of 547 eyes (234 dropless vs. 313 standard) with phacoemulsification cataract surgery and minimum of 1-month follow-up with best-corrected visual acuity (BCVA) were included. Dropless received 40 mg sub-Tenon\u27s triamcinolone and intracameral moxifloxacin. Patients were followed at postoperative day 1 (POD1), week 1 (POW1), and month 1 (POM1). Postoperative rate of BCVA better than 20/40 (Good vision) and rate of complications were compared between groups.
Results  Good vision on POM1 in dropless (77.8%) was noninferior to standard (75.1%, p  = 0.80). Complication rate in dropless (28.6%) was noninferior to standard (24.0%, p  = 0.13). Intraocular pressure (IOP) elevation on POD1 ( p  = 0.041) and anterior chamber (AC) cells on POW1 and POM1 ( p  \u3c 0.001) were more frequent in dropless. Mean spherical equivalent at POM1 was better in dropless (−0.37 D [±0.81 D]) compared with standard (−0.61D [±0.77 D], p  = 0.001). Early posterior capsular opacification (early PCO) was more frequent in dropless ( p  = 0.042).
Conclusions  Postoperative rate of BCVA better than 20/40 and rate of postoperative complications were noninferior, although dropless had higher rates of AC inflammation, IOP elevation, and early PCO
Determining Risk Factors That affect Progression in Patients With Nonproliferative Diabetic Retinopathy
PURPOSE: to determine risk factors that affect nonproliferative diabetic retinopathy (NPDR) progression and establish a predictive model to estimate the probability of and time to progression in NPDR.
RESULTS: Two hundred thirty patients were included. Initially, 65 eyes (28.3%) had no retinopathy; 73 (31.7%) mild NPDR; 60 (26.1%) moderate NPDR; and 32 (13.9%) severe NPDR. Patients were followed for a mean of 5.8 years (±2.0 years; range 2.1-9.4 years). 164 (71.3%) eyes progressed during the follow-up. Time-independent risk factors affecting progression rate were age (hazard ratio (HR) = 0.99,
CONCLUSION: HbA1c level is a significant modifiable risk factor in controlling the progression of DR. The proposed model could be used to predict the time and rate of progression based on an individual\u27s risk factors. A prospective multicenter study should be conducted to further validate our model
Defining the Role of Ab Externo Xen Gel Stent in Glaucomatous Eyes With Prior Failed Surgical Intervention
PURPOSE: To evaluate the safety and efficacy of Xen45 Gel stent (Xen; Allergan) in eyes that have failed prior surgical intervention, compared to traditional glaucoma drainage device (GDD) or continuous-wave cyclophotocoagulation (CPC). Since this population has low expected success rates with additional surgery, it is vital to compare to standard-of-care surgical options.
METHODS: Retrospective, single-center, case-control study of ab externo transconjunctival Xen shunt in eyes that have previously undergone trabeculectomy and/or GDD surgery. Postoperative data were collected for 18 months. Failure was defined as no light perception, additional glaucoma surgery required, or intraocular pressure (IOP) of \u3c 6 mmHg after 6 weeks postoperatively.
RESULTS: Eighteen Xen eyes and 36 control eyes matched on both glaucoma type and previous glaucoma surgeries were included. Seventy-two percent had primary open angle glaucoma, 11% uveitic, 6% primary angle closure, 6% pseudoexfoliation, and 6% pigmentary glaucoma. Fifty-six percent of eyes in each group had prior trabeculectomy, 28% of Xen and 31% of control eyes had prior GDD, and 17% of Xen and 14% of control eyes had both. Baseline medicated IOP was lower in the Xen group (21.8 ± 7.2) compared to controls (27.5 ± 9.4, P = 0.043). The cumulative failure rate at year 1 was 17% for Xen and 20% for controls (P = 0.57). Mean survival time was 14.1 (± 1.5) months and 11.4 (± 0.6) months for controls. There was no difference in minor complication rates between groups (P = 0.65), but the Xen group had a significantly lower rate of serious complications (P = 0.043) defined as vision threatening or requiring surgical intervention in the operating room. When censored for additional glaucoma procedures, there were no differences at year 1 in IOP, change in IOP, number of IOP-lowering medications, or number of medications reduced from baseline.
CONCLUSIONS: The Xen shunt provides a reasonable alternative to current standard of care, with a similar failure rate at year 1, with a noninferior IOP reduction compared to GDD and CPC, and a preferred safety profile
Prognosticators of Visual Acuity After Indirect Traumatic Optic Neuropathy
BACKGROUND: The purpose of this study is to determine whether there are radiographic and systemic clinical characteristics that can predict final visual outcomes in patients with indirect traumatic optic neuropathy (iTON).
METHODS: This study is a retrospective, multicenter case series of adult patients with iTON treated initially at large, urban, and/or academic trauma centers with follow-up at an affiliated ophthalmology clinic. In addition to detailed cranial computed tomography characteristics, demographics, systemic comorbidities, coinjuries, blood products administered, and intracranial pressure, along with other factors, were gathered. LogMAR visual acuity (VA) at the initial presentation to the hospital and up to 12 months follow-up was collected.
RESULTS: Twenty patients met inclusion criteria; 16 (80%) were men with a mean age of 40.9 years (±20.9). Mean initial VA was 1.61 logMAR (∼20/800, ± 0.95), and final VA was 1.31 logMAR (∼20/400, ± 1.06). Three patients (4 eyes) had no light perception (NLP) VA at presentation and remained NLP at final follow-up. Of the predictors analyzed, only the initial VA was found to be a significant predictor of visual outcome. The presence of orbital fractures, intraconal and/or extraconal hemorrhage, as well as systemic comorbidities, were not found to significantly affect visual outcome.
CONCLUSIONS: After evaluating multiple factors, initial VA was the only factor associated with visual prognosis in iTON. This knowledge may better enable clinicians to predict visual prognosis and set reasonable expectations with patients and families at the time of injury
Direct Costs of Second Aqueous Shunt Implant Versus Transscleral Cyclophotocoagulation (The Assists Trial)
PRCIS: The cost of cyclophotocoagulation is less than the cost of a second glaucoma drainage device.
PURPOSE: To compare the total direct costs of implantation of a second glaucoma drainage device (SGDD) with transscleral cyclophotocoagulation (CPC) for patients with inadequately controlled intraocular pressure (IOP) reduction, despite the presence of a preexisting glaucoma drainage device in the ASSISTS clinical trial.
METHODS: We compared the total direct cost per patient, including the initial study procedure, medications, additional procedures, and clinic visits during the study period. The relative costs for each procedure during the 90-day global period and the entire study period were compared. The cost of the procedure, including facility fees and anesthesia costs, were determined using the 2021 Medicare fee schedule. Average wholesale prices for self-administered medications were obtained from AmerisourceBergen.com. The Wilcoxon rank sum test was used to compare costs between procedures.
RESULTS: Forty-two eyes of 42 participants were randomized to SGDD (n=22) or CPC (n=20). One CPC eye was lost to follow-up after initial treatment and was excluded. The mean (±SD, median) duration of follow-up was 17.1 (±12.8, 11.7) months and 20.3 (±11.4, 15.1) months for SGDD and CPC, respectively ( P =0.42, 2 sample t test). The mean total direct costs (±SD, median) per patient during the study period were 3421, 4090 (±3566) for the CPC group ( P
CONCLUSION: The total direct cost in the SGDD group was more than double that in the CPC group, driven largely by the cost of the study procedure. The costs of IOP-lowering medications were not significantly different between groups. When considering treatment options for patients with a failed primary GDD, clinicians should be aware of differences in costs between these treatment strategies
Synaptic Origins of the Complex Receptive Field Structure in Primate Smooth Monostratified Retinal Ganglion Cells
Considerable progress has been made in studying the receptive fields of the most common primate retinal ganglion cell (RGC) types, such as parasol RGCs. Much less is known about the rarer primate RGC types and the circuitry that gives rise to noncanonical receptive field structures. The goal of this study was to analyze synaptic inputs to smooth monostratified RGCs to determine the origins of their complex spatial receptive fields, which contain isolated regions of high sensitivity called hotspots. Interestingly, smooth monostratified RGCs co-stratify with the well-studied parasol RGCs and are thus constrained to receiving input from bipolar and amacrine cells with processes sharing the same layer, raising the question of how their functional differences originate. Through 3D reconstructions of circuitry and synapses onto ON smooth monostratified and ON parasol RGCs from central macaque retina, we identified four distinct sampling strategies employed by smooth and parasol RGCs to extract diverse response properties from co-stratifying bipolar and amacrine cells. The two RGC types differed in the proportion of amacrine cell input, relative contributions of co-stratifying bipolar cell types, amount of synaptic input per bipolar cell, and spatial distribution of bipolar cell synapses. Our results indicate that the smooth RGC\u27s complex receptive field structure arises through spatial asymmetries in excitatory bipolar cell input which formed several discrete clusters comparable with physiologically measured hotspots. Taken together, our results demonstrate how the striking differences between ON parasol and ON smooth monostratified RGCs arise from distinct strategies for sampling a common set of synaptic inputs
Direct Costs of Second Aqueous Shunt Implant Versus Transscleral Cyclophotocoagulation (The Assists Trial)
PRCIS: The cost of cyclophotocoagulation is less than the cost of a second glaucoma drainage device.
PURPOSE: To compare the total direct costs of implantation of a second glaucoma drainage device (SGDD) with transscleral cyclophotocoagulation (CPC) for patients with inadequately controlled intraocular pressure (IOP) reduction, despite the presence of a preexisting glaucoma drainage device in the ASSISTS clinical trial.
METHODS: We compared the total direct cost per patient, including the initial study procedure, medications, additional procedures, and clinic visits during the study period. The relative costs for each procedure during the 90-day global period and the entire study period were compared. The cost of the procedure, including facility fees and anesthesia costs, were determined using the 2021 Medicare fee schedule. Average wholesale prices for self-administered medications were obtained from AmerisourceBergen.com. The Wilcoxon rank sum test was used to compare costs between procedures.
RESULTS: Forty-two eyes of 42 participants were randomized to SGDD (n=22) or CPC (n=20). One CPC eye was lost to follow-up after initial treatment and was excluded. The mean (±SD, median) duration of follow-up was 17.1 (±12.8, 11.7) months and 20.3 (±11.4, 15.1) months for SGDD and CPC, respectively ( P =0.42, 2 sample t test). The mean total direct costs (±SD, median) per patient during the study period were 3421, 4090 (±3566) for the CPC group ( P
CONCLUSION: The total direct cost in the SGDD group was more than double that in the CPC group, driven largely by the cost of the study procedure. The costs of IOP-lowering medications were not significantly different between groups. When considering treatment options for patients with a failed primary GDD, clinicians should be aware of differences in costs between these treatment strategies
Risk Factors for Tube Shunt Exposure: A Matched Case-Control Study
Purpose. To evaluate potential risk factors for developing tube shunt exposure in glaucoma patients. Patients and Methods. Forty-one cases from 41 patients that had tube shunt exposure from 1996 to 2005 were identified from the Robert Cizik Eye Clinic and Bascom Palmer Eye Institute. Each case was matched with 2 controls of the same gender and with tube shunts implanted within 6 months of the index case. Conditional logistic regression was used to determine risk factors. Results. The study cohort includes a total of 121 eyes from 121 patients. The mean age was 63.6 ± 19.7 years, ranging from 1 to 96 years. The average time to exposure was 19.29 ± 23.75 months (range 0.36–85.74 months). Risk factors associated with tube exposure were Hispanic ethnicity (; OR = 3.6; 95% CI, 1.3–9.7), neovascular glaucoma (; OR = 28.5; 95% CI, 2.6–316.9), previous trabeculectomy (; OR = 5.3; 95% CI, 1.6–17.7), and combined surgery (; OR = 3.7; 95% CI, 1.1–12.7). Conclusions. Hispanic ethnicity, neovascular glaucoma, previous trabeculectomy, and combined surgery were identified as potential risk factors for tube shunt exposure. These potential risk factors should be considered when determining the indication for performing tube shunt implantation and the frequency of long-term followup
- …