7 research outputs found

    Fully Implantable Hearing Aids

    No full text

    Stapes Surgery for Otosclerosis in Patients Presenting with Mixed Hearing Loss

    Full text link
    Objectives: To evaluate the efficacy of stapes surgery in patients presenting with a preoperative mixed hearing loss (bone conduction thresholds ≥40 dB; 40 < air conduction thresholds (AC) < 85 dB). Patients and methods: A total of 30 patients (32 ears) with mixed hearing loss who underwent primary stapedotomy were evaluated. Audiometric parameters were assessed before and after surgery. Contralateral thresholds were also reported. The need for a hearing aid (HA) after surgery and its impact on quality of life were also measured. Results: AC and word recognition at 40, 55 and 70 dB were significantly improved after stapes surgery. Only 16.6% of the patients needed an HA after surgery and reported being satisfied with the aid. Conclusion: Stapes surgery improved auditory function in patients with mixed hearing loss, allowing most patients to delay the need for an HA without worsening their quality of life.Peer reviewe

    Hearing Preservation in Cochlear Implantation and Drug Treatment.

    Full text link
    Insertion of an electrode array into the cochlea produces immediate damage to the inner ear, which is responsible for a hearing loss. In addition, a delayed hearing loss can be observed. In order to maximize hearing preservation after insertion of an electrode and to enhance the performance of the cochlear implant, it has been proposed to deliver pharmacological agents to the inner ear. Molecules can be administered locally to the inner ear through a direct perilymphatic perfusion or through the round window membrane. These modalities of treatment have already been successfully applied to some patients with inner ear diseases. In this paper, we will review some basic aspects of drug delivery to the inner ear to prevent the degeneration of the neurosensory hair cells and auditory neurons, and the actual applicability to humans in order to maintain hearing function after the insertion of electrodes of a cochlear implant

    The Use of a Robot to Insert an Electrode Array of Cochlear Implants in the Cochlea: A Feasibility Study and Preliminary Results.

    Full text link
    peer reviewed[en] INTRODUCTION: Cochlear implants (CIs) are commonly used for the rehabilitation of profound bilateral hearing loss. However, patients with substantial residual acoustic hearing are potential CI candidates. Because of both improvements in technology and advancements in surgical techniques, it may be possible to preserve hearing to some extent. For more than a decade, it has been suggested that robots are used to perform middle ear surgery. We evaluated the use of the RobOtol® otologic robot specifically to insert CI electrodes into the inner ear. METHODS: CI surgery with the conventional approach was performed under general anesthesia. The MED-El Flex 24-electrode array was inserted using RobOtol®. Video recordings were used to calculate the speed of insertion. The positions of the electrodes were evaluated using a cone beam CT. All subjects underwent pure-tone audiometry tests before and after surgery, and the pure-tone average (PTA) was calculated from 250 to 4,000 Hz. RESULTS: The robot inserted implants in 5 patients, and complete insertion of the electrode array was achieved. The speed of insertion of the electrode array was 0.88 ± 0.12 mm/s. The mean loss of the PTA for 5 frequencies (250, 500, 1,000, 2,000, and 4,000 Hz) was 13.60 ± 7.70 dB. Only 1 patient showed a loss of the PTA by >20 dB. For these 5 patients, the cone beam CT findings showed that all the electrode arrays were in the tympanic ramp and had a grade of 0. The results were compared with those obtained from a cohort of 17 patients who underwent manual implantation of a MED-El Flex 24-electrode array. CONCLUSION: To minimize disturbance to the cochlea while atraumatic electrode arrays are inserted, electrodes can be inserted at a constant, slow speed in the inner ear with the assistance of the RobOtol® robot in a normal clinical surgical setting

    Audiometric results after stapedotomy operations in patients with otosclerosis and preoperative small air-bone gaps

    Full text link
    Objectives: The efficacy of stapedotomies performed on patients with small air-bone gaps (<25 dB, sABG) was compared to the efficacy of the operation in patients who had otosclerosis with high air-bone gaps (≥25 dB, hABG). Methods: This retrospective study evaluates the short-term postoperative air and bone conduction thresholds and air-bone gaps after 181 CO2 laser stapedotomies. Results: A significantly smaller air-bone gap (ABG) and lower air conduction thresholds after surgery were observed in the group of patients who underwent surgery with preoperative ABGs of less than 25 dB. Bone conduction thresholds improve in sABG group after surgery. Conclusions: The results after stapedotomies are good even if the preoperative air-bone gap is small and that the overall risk of hearing deterioration due to stapes surgery remains low
    corecore