    Conductive heating mattress leads to ECG changes that mimic pacemaker spikes

    No full text
    Hypothermia is a common perioperative complication. To prevent perioperative hypothermia amongst other things electrical heating mattresses are used. We have made an observation with the use of an electrical heating mattress that may confuse users. In this case the ECG monitoring suddenly showed spikes that looked like spikes from an implanted pacemaker. When turning off the heating mattress the spikes disappeared and returned after turning on the heating mattress again

    Intraoperative zero-heat-flux thermometry overestimates esophageal temperature by 0.26 °C: an observational study in 100 infants and young children

    No full text
    In pediatric anesthesia, deviations from normothermia can lead to many complications, with infants and young children at the highest risk. A measurement method for core temperature must be clinically accurate, precise and should be minimally invasive. Zero-heat-flux (ZHF) temperature measurements have been evaluated in several studies in adults. We assessed the agreement between the 3M Bair Hugge