2 research outputs found

    Instrumentation and Automated Photometric Titration Procedure for Total Acidity Determination in Red Wine Employing a Multicommuted Flow System

    Get PDF
    An automated procedure for photometric titration of red wine and associated instrumentation is described. The procedure was based on the flow-batch approach implemented employing multicommutation. The photometric detection was carried out using a homemade LED-based photometer. The mixing device, LED, and photodetector were attached to the titration chamber in order to form a compact and small-sized unit. The flow system comprised an automatic injector and three-way solenoid valves, which were controlled by a microcomputer through an electronic interface card. The software, written in Quick BASIC 4.5, was designed with abilities to accomplish all steps of the titration procedure including data acquisition and real-time processing to decide about the course of titration in the following step and so forth, until the titration endpoint was reached. The usefulness of the proposed titration system was demonstrated by analyzing red wine samples. When results were compared with those obtained using the AOAC reference method, no significant difference was observed at the [Formula: see text] confidence level. A relative standard deviation of ca [Formula: see text] ([Formula: see text]) was obtained when processing a typical red wine sample containing 7.3 [Formula: see text] total acidity expressed as tartaric acid