    Fast Estimation of the Vascular Cooling in RFA Based on Numerical Simulation

    Get PDF
    We present a novel technique to predict the outcome of an RF ablation, including the vascular cooling effect. The main idea is to separate the problem into a patient independent part, which has to be performed only once for every applicator model and generator setting, and a patient dependent part, which can be performed very fast. The patient independent part fills a look-up table of the cooling effects of blood vessels, depending on the vessel radius and the distance of the RF applicator from the vessel, using a numerical simulation of the ablation process. The patient dependent part, on the other hand, only consists of a number of table look-up processes. The paper presents this main idea, along with the required steps for its implementation. First results of the computation and the related ex-vivo evaluation are presented and discussed. The paper concludes with future extensions and improvements of the approach

    Parameter Estimation for Personalization of Liver Tumor Radiofrequency Ablation

    Get PDF
    International audienceMathematical modeling has the potential to assist radiofrequency ablation (RFA) of tumors as it enables prediction of the extent of ablation. However, the accuracy of the simulation is challenged by the material properties since they are patient-specific, temperature and space dependent. In this paper, we present a framework for patient specific radiofrequency ablation modeling of multiple lesions in the case of metastatic diseases. The proposed forward model is based upon a computational model of heat diffusion, cellular necrosis and blood flow through vessels and liver which relies on patient images. We estimate the most sensitive material parameters, those need to be personalized from the available clinical imaging and data. The selected parameters are then estimated using inverse modeling such that the point to-mesh distance between the computed necrotic area and observed lesions is minimized. Based on the personalized parameters, the ablation of the remaining lesions are predicted. The framework is applied to a dataset of seven lesions from three patients including pre- and post-operative CT images. In each case, the parameters were estimated on one tumor and RFA is simulated on the other tumor(s) using these personalized parameters, assuming the parameters to be spatially invariant within the same patient. Results showed significantly good correlation between predicted and actual ablation extent (average point-to-mesh errors of 4.03 mm)

    Radiofrequency ablation planning beyond simulation

    No full text
    It is a challenging task to plan a radiofrequency (RF) ablation therapy to achieve the best outcome of the treatment and avoid recurrences at the same time. A patient specific simulation in advance that takes the cooling effect of blood vessels into account is a helpful tool for radiologists, but this needs a very high accuracy and thus high computational costs. In this work, we present various methods, which improve and extend the planning of an RF ablation procedure. First, we discuss two extensions of the simulation model to obtain a higher accuracy, including the vaporization of the water in the tissue and identifying the model parameters and to analyze their uncertainty. Furthermore, we discuss an extension of the planning procedure namely the optimization of the probe placement, which optimizes the overlap of the tumor area with the estimated coagulation in order to avoid recurrences. Since the optimization is constrained by the model, we have to take into account the uncertainties in the model parameters for the optimization as well. Finally, applications of our methods to a real RF ablation case are presented
    corecore