Polymeric Sorbent with Controlled Surface Polarity: An Alternate for Solid-Phase Extraction of Nerve Agents and Their Markers from Organic Matrix
Extraction and identification
of lethal nerve agents and their
markers in complex organic background have a prime importance from
the forensic and verification viewpoint of the Chemical Weapons Convention
(CWC). Liquid–liquid extraction with acetonitrile and commercially
available solid phase silica cartridges are extensively used for this
purpose. Silica cartridges exhibit limited applicability for relatively
polar analytes, and acetonitrile extraction shows limited efficacy
toward relatively nonpolar analytes. The present study describes the
synthesis of polymeric sorbents with tunable surface polarity, their
application as a solid-phase extraction (SPE) material against nerve
agents and their polar as well as nonpolar markers from nonpolar organic
matrices. In comparison with the acetonitrile extraction and commercial
silica cartridges, the new sorbent showed better extraction efficiency
toward analytes of varying polarity. The extraction parameters were
optimized for the proposed method, which included ethyl acetate as
an extraction solvent and <i>n</i>-hexane as a washing solvent.
Under optimized conditions, method linearity ranged from 0.10 to 10
μg mL<sup>–1</sup> (<i>r</i><sup>2</sup> =
0.9327–0.9988) for organophosphorus esters and 0.05–20
μg mL<sup>–1</sup> (<i>r</i><sup>2</sup> =
0.9976–0.9991) for nerve agents. Limits of detection (S:N =
3:1) in the SIM mode were found in the range of 0.03–0.075
μg mL<sup>–1</sup> for organophosphorus esters and 0.015–0.025
μg mL<sup>–1</sup> for nerve agents. Limits of quantification
(S:N = 10:1) were found in the range of 0.100–0.25 μg
mL<sup>–1</sup> for organophosphorus esters and 0.05–0.100
μg mL<sup>–1</sup> for nerve agents in the SIM mode.
The recoveries of the nerve agents and their markers ranged from 90.0
to 98.0% and 75.0 to 95.0% respectively. The repeatability and reproducibility
(with relative standard deviations (RSDs) %) for organophosphorus
esters were found in the range of 1.35–8.61% and 2.30–9.25%
respectively. For nerve agents, the repeatability range from 1.00
to 7.75% and reproducibility were found in the range of 2.17–6.90%