10 research outputs found

    Prospective study of toric IOL outcomes based on the Lenstar LS 900® dual zone automated keratometer

    Get PDF
    <p>Abstract</p> <p>Background</p> <p>To establish clinical expectations when using the Lenstar LS 900® dual-zone automated keratometer for surgery planning of toric intraocular lenses.</p> <p>Methods</p> <p>Fifty eyes were measured with the Lenstar LS 900® dual-zone automated keratometer . Surgical planning was performed with the data from this device and the known surgically induced astigmatism of the surgeon. Post-operative refractions and visual acuity were measured at 1 month and 3 months.</p> <p>Results</p> <p>Clinical outcomes from 43 uncomplicated surgeries showed an average post-operative refractive astigmatism of 0.44D ±0.25D. Over 70% of eyes had 0.50D or less of refractive astigmatism and no eye had more than 1.0D of refractive astigmatism. Uncorrected visual acuity was 20/32 or better in all eyes at 3 months, with 70% of eyes 20/20 or better. A significantly higher number of eyes had 0.75D or more of post-operative refractive astigmatism when the standard deviation of the pre-operative calculated corneal astigmatism angle, reported by the keratometer, was > 5 degrees.</p> <p>Conclusions</p> <p>In this single-site study investigating the use of the keratometry from the Lenstar LS 900® for toric IOL surgical planning, clinical outcomes appear equivalent to those reported in the literature for manual keratometry and somewhat better than has been reported for some previous automated instruments. A high standard deviation in the pre-operative calculated astigmatism angle, as reported by the keratometer, appears to increase the likelihood of higher post-operative refractive astigmatism.</p
    corecore