Clinical results of entrance dose in vivo dosimetry for high energy photons in external beam radiotherapy using MOSFETs
Copyright © 2007 ACPSEM. All rights reserved. The document attached has been archived with permission from the publisher.Thomson and Nielsen TN-502 RD MOSFETs were used for entrance dose in vivo dosimetry for 6 and 10 MV photons. A total of 24 patients were tested, 10 breast, 8 prostate, 5 lung and 1 head and neck. For prostates three fields were checked. For all other plans all fields were checked. An action threshold of 8% was set for any one field and 5% for all fields combined. The total number of fields tested was 56, with a mean discrepancy of 1.4% and S.D. of 2.6%. Breasts had a mean discrepancy of 1.8% and S.D. of 2.8%. Prostates had a mean discrepancy of 1.3% and S.D. of 2.9%. For 3 fields combined, prostates had a mean of 1.3% and S.D. of 1.8%. These results are similar to results obtained with diodes and TLDs for the same techniques.J. P. Morton, M. Bhat, T. Williams and A. Kovend
Evaluation of MOSFETs for entrance dose dosimetry for 6 and 10 MV photons with a custom made build up cap
Copyright © 2007 ACPSEM. All rights reserved. The dcoument attached has been archived with permission from the publisher.Commercially available MOSFETs, Thomson and Nielsen TN502-RD, were evaluated for suitability as an entrance dose in vivo dosimeter for 6MV and 10MV. Detector response was normally distributed around a mean (skewness=-0.01±0.24, kurtosis=-0.09±0.48) with a mean of 110.6 mV/Gy, with a standard deviation of 2.4% at 0.86 Gy. The standard deviation of readings increased with decreasing dose and increased at a rate greater than inverse square. The linearity coefficient was 0.9999. No significant dependence on angle, field size, dose rate, energy or time was observed. As such, they would be useful for entrance dose in vivo dosimetry. With a custom made build up cap corrections were required for field size, wedge, beam energy and tray factors, showing that build up cap design is an important consideration for entrance dose in vivo dosimetry using MOSFETs.J. P. Morton, M. Bhat, A. Kovendy and T. Williamshttp://www.acpsem.org.au/journal/abstract/abstract_3002.html#abs0