Traceability of the 10 kΩ Standard at IEN
none6The traditional scaling method used at the Istituto Elettrotecnico Nazionale (IEN) for the calibration of the 10-k Omega standard is compared with more straightforward techniques: one based on the linearity of a DVM and the other on a commercial current comparator bridge. For the three methods, the measurement results and the uncertainty budgets are reported. The agreement is better than 1 x 10(-7).mixedBoella G; Capra PP; Cassiago C; Cerri R; Reedtz GM; Sosso ABoella, G; Capra, PIER PAOLO; Cassiago, Cristina; Cerri, R; Reedtz, Gm; Sosso, Andre