1 research outputs found

    Positive Contrast Susceptibility MR Imaging Using GPU-based Primal-Dual Algorithm

    Full text link
    The susceptibility-based positive contrast MR technique was applied to estimate arbitrary magnetic susceptibility distributions of the metallic devices using a kernel deconvolution algorithm with a regularized L-1 minimization.Previously, the first-order primal-dual (PD) algorithm could provide a faster reconstruction time to solve the L-1 minimization, compared with other methods. Here, we propose to accelerate the PD algorithm of the positive contrast image using the multi-core multi-thread feature of graphics processor units (GPUs). The some experimental results showed that the GPU-based PD algorithm could achieve comparable accuracy of the metallic interventional devices in positive contrast imaging with less computational time. And the GPU-based PD approach was 4~15 times faster than the previous CPU-based scheme.Comment: 4 pages, 6 figures, Accepted at the 42nd Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC 2020