    Finite element analysis of gradient coil deformation and vibration in NMR microscopy

    Get PDF
    Resolution degradation due to gradient coil deformation and vibration in NMR microscopy is investigated using finite element analysis. From the analysis, deformations due to the Lorentz force can be as large as 1-10 ÎĽm depending on the gradient strength and coil frame material. Thus, these deformations can be one of the major resolution limiting factors in NMR microscopy. Coil vibration, which depends on the input current waveform and resolution degradation due to time-variant deformation and time-invariant deformation are investigated by numerical simulations