Achieving grazing-incidence ultra-small-angle X-ray scattering in a laboratory setup

Abstract

A grazing-incidence sample stage was designed for realizing grazing-incidence scattering measurements, especially in the ultra-small-angle regime, in a modified Xenocs Xeuss system in the laboratory. The designed sample stage, which is composed of four separate motorized positioning stages, allows the sample to be moved along four different directions to locate it in the proper position for scattering measurement. In an effort to realize grazing-incidence ultra-small-angle scattering (GIUSAXS) measurements, both the separation of the collimation slit systems and the sample-to-detector distance have been lengthened. At a separation of the collimation slit systems of 2400mm and a sample-to-detector distance of 6558mm, the effective smallest scattering vector magnitude q(min) reaches 0.01nm(-1). A colloidal crystalline thin film obtained from drying a polystyrene latex dispersion on silicon substrate was measured in the setup in GIUSAXS mode at different beam sizes. The resultant GIUSAXS patterns at smaller beam sizes reveal fine crystalline structures in the film

Similar works

Full text

thumbnail-image

Changchun Institute of Applied Chemistry, Chinese Academy Of Sciences

redirect
Last time updated on 13/03/2018