We have developed a simple and selective method for the electrochemical detection of hydrazine (HZ) using poly(dopamine) (pDA)-modified indium tin oxide (ITO) electrodes. Modification with pDA was easily achieved by submerging the ITO electrode in a DA solution for 30 min. The electrocatalytic oxidation of HZ on the pDA-modified ITO electrode was measured by cyclic voltammetry. In buffer solution, the concentration range for linear HZ detection was 100 µM–10 mM, and the detection limit was 1 µM. The proposed method was finally used to determine HZ in tap water to simulate the analysis of real samples. This method showed good recovery (94%–115%) and was not affected by the other species present in the tap water samples