Article thumbnail

The flow rate of the elastomeric balloon infusor is influenced by the internal pressure of the infusor.

By I. S. Chung, H. S. Cho, J. A. Kim and K. H. Lee

Abstract

The purpose of this study was to evaluate the flow rate of elastomeric balloon infusor composed of rubber or silicone materials. Two models were studied: the Baxter Twoday Infusor and Advance Silicone Infusor. Each infusion device has a preset flow rate of 2 mL/hr. Fifteen units of each device were filled with 100 mL of normal saline. The flow rate was measured gravimetically using an electronic balance. The internal pressure of the balloon infusor was measured with a Digital pressure meter via the pressure transducer. The monitored internal pressure of the two kinds of infusor was not maintained uniformly during the entire delivery period, which was divided into 3 phases: the phase of decreasing, maintained, and increasing pressure during the delivery of drug. Both devices initially infused at a relatively high rate, followed by a somewhat steady flow rate. The flow rate distinctly increased in the small residual volume. The flow rate of the balloon infusor used in this study was not sustained uniformly during the entire delivery period and was in proportion to the internal pressure of the infusor regardless of the materials

Topics: Research Article
Publisher: Korean Academy of Medical Sciences
OAI identifier: oai:pubmedcentral.nih.gov:3054806
Provided by: PubMed Central

To submit an update or takedown request for this paper, please submit an Update/Correction/Removal Request.

Suggested articles

Citations

  1. (1986). A disposable device for infusion analgesia: an evaluation of the Travenol infusor. Anaesthesia
  2. (1989). A disposable device for patient-controlled analgesia with fentanyl. Anaesthesia
  3. (1992). An automated system for testing the accuracy of patient-controlled analgesia devices. Anaesthesia
  4. (1997). Comparison of accuracy and cost of disposable, nonmechanical pumps used for epidural infusions. Reg Anest
  5. (1992). Electronic and disposable patient-controlled analgesia systems: a comparison of the Graseby and Baxter systems after major gynaecological surgery. Anaesthesia
  6. (1987). Evaluation of a disposable non-electric, patient-controlled analgesia device for post operative pain. Clin Pharm
  7. (1992). Evaluation of a disposable patient-controlled analgesia device in children.
  8. (1992). Evaluation of a disposable, elastomeric infusion device in the home environment.
  9. (1995). Evaluation of flow rates for six disposable infusion devices.
  10. (1993). Laboratory evaluation of the Baxter patient-controlled analgesia infusor system: a disposable patient-controlled analgesia device. Anesth Analg
  11. (1989). Patient-controlled analgesia: a review of effectiveness of therapy and an evaluation of currently available devices. Drug Intell Clin Pharm
  12. (1992). Patient-controlled analgesia. A serious incident. Anaesthesia
  13. (1994). Prolonged infusion times with disposable elastomeric infusion devices.
  14. (1990). Simplified postoperative patient-controlled analgesia on a gynecologic oncology service. Gynecol Oncol
  15. (1997). The performance of three portable infusion-pump devices set to deliver 2 mL/hr.