Integrated microanalytical technology enabling rapid and automated protein identification

Abstract

Protein identification through peptide mass mapping by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry (MALDI-TOF MS) has become a standard technique, used in many laboratories around the world. The traditional methodology often includes long incubations (6-24 h) and extensive manual steps. In an effort to address this, an integrated microanalytical platform has been developed for automated identification of proteins. The silicon micromachined analytical tools, i.e., the microchip immobilized enzyme reactor (μ-chip IMER), the piezoelectric microdispenser, and the high-density nanovial target plates, are the cornerstones in the system. The μ-chip IMER provides on-line enzymatic digestion of protein samples (1 μL) within 1-3 min, and the microdispenser enables subsequent on- line picoliter sample preparation in a high-density format. Interfaced to automated MALDI-TOF MS, these tools compose a highly efficient platform that can analyze 100 protein samples in 3.5 h. Kinetic studies on the microreactors are reported as well as the operation of this microanalytical platform for protein identification, wherein lysozyme, myoglobin, ribonuclease A, and cytochrome c have been identified with a high sequence coverage (50-100%)

Similar works

Full text

thumbnail-image

Lund University Publications

redirect
Last time updated on 18/06/2017

This paper was published in Lund University Publications.

Having an issue?

Is data on this page outdated, violates copyrights or anything else? Report the problem now and we will take corresponding actions after reviewing your request.