The American Association for Thoracic Surgery. Published by Mosby, Inc.
Doi
Abstract
ObjectiveThis prospective randomized study compares a porcine with a bovine bioprosthesis in the aortic position with regard to hemodynamic performance during exercise.MethodsBetween August of 2000 and December of 2002, 136 patients underwent aortic valve replacement with the porcine Medtronic Mosaic (n = 66) or the bovine Carpentier-Edwards Perimount (n = 70) bioprosthesis. Transthoracic echocardiography was performed to assess hemodynamic and dimensional data preoperatively and 10 months postoperatively; the latter follow-up included stress echocardiography with treadmill exercise.ResultsAt rest and during exercise (25 and 50 W), there was a significant difference in mean pressure gradient between the bovine and the porcine valves with labeled sizes 21 and 23, with superiority of the Perimount prosthesis. There was no difference in effective orifice area and incidence of patient-prosthesis mismatch among all sizes. The left ventricular mass index decreased significantly within 10 months postoperatively in the size 23 bovine group and the size 25 porcine group.ConclusionsOur data show a significant superiority of pressure gradients for the bovine bioprosthesis, especially with small valve sizes, when compared with the porcine device, which is more distinctive during exercise