Development and validation of a liquid chromatography-tandem mass spectrometry assay for the simultaneous quantitation of microcystin-RR and its metabolites in fish liver

Abstract

A novel method for identification and quantification of microcystin-RR (MC-RR) and its metabolites (MC-RR-GSH and MC-RR-Cys) in the fish liver was developed and validated. These analytes were simultaneously extracted from fish liver using water containing EDTA with 5% acetic acid, followed by a mixed-mode cation-exchange SPE (Oasis MCX) and subsequently determined by liquid chromatography-electrospray ionization ion trap mass spectrometry (LC-ESI-ITMS). Extraction parameters including volume and pH of eluting solvents, were optimized. Best recoveries were obtained by using 10 mL of 15% ammonia solution in methanol. The mean recoveries at three concentrations (0.2, 1.0, and 5.0 mu g g(-1) dry weight [DW]) for MC-RR, MC-RR-GSH and MC-RR-Cys were 93.6-99%,68.1-73.6% and 90.0-95.2%, respectively. Method detection limit (MDL) were 4, 7 and 5 ng g(-1) DW for MC-RR, MC-RR-GSH and MC-RR-Cys, respectively. Limits of quantification (LOQs) for MC-RR, MC-RR-GSH and MC-RR-Cys were calculated to be 10, 18 and 13 ng g(-1) DW, respectively. Finally, this method was successfully applied to the identification and quantification of MC-RR, MC-RR-GSH and MC-RR-Cys in the liver of bighead carp with acute exposure of MCs. (C) 2009 Elsevier B.V. All rights reserved

Similar works

Full text

thumbnail-image

Institute of Hydrobiology, Chinese Academy Of Sciences

redirect
Last time updated on 15/03/2018

Having an issue?

Is data on this page outdated, violates copyrights or anything else? Report the problem now and we will take corresponding actions after reviewing your request.