Capillary Zone Electrophoresis–Electrospray
Ionization Tandem Mass Spectrometry System with an Immobilized Trypsin
Microreactor for Online Digestion and Analysis of Picogram Amounts
of RAW 264.7 Cell Lysate
- Publication date
- Publisher
Abstract
A capillary zone electrophoresis
(CZE) electrospray ionization
(ESI) tandem mass spectrometry (MS/MS) system was integrated with
an immobilized trypsin microreactor. The system was evaluated and
then applied for online digestion and analysis of picogram loadings
of RAW 264.7 cell lysate. Protein samples were dissolved in a buffer
containing 50% (v/v) acetonitrile (ACN), and then directly loaded
into the capillary for digestion, followed by CZE separation and MS/MS
identification. The organic solvent (50% (v/v) ACN) assisted the immobilized
trypsin digestion and simplified the protein sample preparation protocol.
Neither protein reduction nor alkylation steps were employed, which
minimized sample loss and contamination. The integrated CZE–ESI-MS/MS
system generated confident identification of bovine serum albumin
(BSA) with 19% sequence coverage and 14 peptide identifications (IDs)
when 20 fmol was loaded. When only 1 fmol of BSA was injected, one
BSA peptide was consistently detected. For the analysis of a standard
protein mixture, the integrated system produced efficient protein
digestion and confident identification for proteins with different
molecular weights and isoelectric points when a low-femtomole amount
was loaded for each protein. We further applied the system for triplicate
analysis of a RAW 264.7 cell lysate; 2 ± 1 and 7 ± 2 protein
groups were confidently identified from only 300 pg and 3 ng loadings,
respectively. The 300 pg sample loading corresponds to the protein
content of three RAW 264.7 cells. In addition to high-sensitivity
analysis, the integrated CZE–ESI-MS/MS system produces good
reproducibility in terms of peptide and protein IDs, peptide migration
time, and peptide intensity