Miniscale Liquid–Liquid Extraction Coupled
with Full Evaporation Dynamic Headspace Extraction for the Gas Chromatography/Mass
Spectrometric Analysis of Polycyclic Aromatic Hydrocarbons with 4000-to-14 000-fold
Enrichment
A new sample preparation approach
of combining a miniscale version
of liquid–liquid extraction (LLE), termed miniscale-LLE (msLLE),
with automated full evaporation dynamic headspace extraction (FEDHS)
was developed. Its applicability was demonstrated in the extraction
of several polycyclic aromatic hydrocarbons (PAHs) (acenaphthylene,
acenaphthene, fluorene, phenanthrene, anthracene, fluoranthene, and
pyrene) from aqueous samples. In the first step, msLLE was conducted
with 1.75 mL of n-hexane, and all of the extract
was vaporized through a Tenax TA sorbent tube via a nitrogen gas flow,
in the FEDHS step. Due to the stronger π–π interaction
between the Tenax TA polymer and PAHs, only the latter, and not n-hexane, was adsorbed by the sorbent. This selectivity
by the Tenax TA polymer allowed an effective concentration of PAHs
while eliminating n-hexane by the FEDHS process.
After that, thermal desorption was applied to the PAHs to channel
them into a gas chromatography/mass spectrometric (GC/MS) system for
analysis. Experimental parameters affecting msLLE (solvent volume
and mixing duration) and FEDHS (temperature and duration) were optimized.
The obtained results achieved low limits of detection (1.85–3.63
ng/L) with good linearity (r2 > 0.9989)
and high enrichment factors ranging from 4200 to 14 100. The
optimized settings were applied to the analysis of canal water sampled
from an industrial area and tap water, and this methodology was compared
to stir-bar sorptive extraction (SBSE). This innovative combined extraction–concentration
approach proved to be fast, effective, and efficient in determining
low concentrations of PAHs in aqueous samples