We describe an AC method for the measurement of the longitudinal (Sxx) and
transverse (Sxy, i.e. Nernst) thermopower of mm-size single crystal samples at
low temperatures (T30 T). A low-frequency (33
mHz) heating method is used to increase the resolution, and to determine the
temperature gradient reliably in high magnetic fields. Samples are mounted
between two thermal blocks which are heated by a sinusoidal frequency f0 with a
p/2 phase difference. The phase difference between two heater currents gives a
temperature gradient at 2f0. The corresponding thermopower and Nernst effect
signals are extracted by using a digital signal processing method due. An
important component of the method involves a superconducting link, YBa2Cu3O7+d
(YBCO), which is mounted in parallel with sample to remove the background
magnetothermopower of the lead wires. The method is demonstrated for the quasi
two-dimensional organic conductor a-(BEDT-TTF)2KHg(SCN)4, which exhibits a
complex, magnetic field dependent ground state above 22.5 T at low
temperatures.Comment: 11 pages, 6 figures, 15 reference