Development of the UV Spectrophotometric Method of Azithromycin in API and Stress Degradation Studies

Abstract

ABSTRACT: Azithromycin (AZI) is a semi-synthetic macrolide antibiotic drug, effective against a wide variety of bacteria. The present study describes a simple, accurate, reproducible and precise UV Spectrophotometric method for the estimation of AZI (pH 6.8 Phosphate buffer). The absorbance maximum (λ max ) for AZI was found to be 208nm. The method reveals high sensitivity, with linearity in the 10 µg/ml to 50 µg/ml range. The lower limit of detection was found to be 1.6µg/ml and the limit of quantification was found to be 5µg/ml. All the calibration curves demonstrated a linear relationship between the absorbance and concentration, with the correlation coefficient higher than 0.99. The % recovery was found to be 99.72%. AZI was also subjected to stress degradation under different conditions recommended by the International Conference on Harmonization (ICH). Introduction