1,441 research outputs found
Accurate calibration of test mass displacement in the LIGO interferometers
We describe three fundamentally different methods we have applied to
calibrate the test mass displacement actuators to search for systematic errors
in the calibration of the LIGO gravitational-wave detectors. The actuation
frequencies tested range from 90 Hz to 1 kHz and the actuation amplitudes range
from 1e-6 m to 1e-18 m. For each of the four test mass actuators measured, the
weighted mean coefficient over all frequencies for each technique deviates from
the average actuation coefficient for all three techniques by less than 4%.
This result indicates that systematic errors in the calibration of the
responses of the LIGO detectors to differential length variations are within
the stated uncertainties.Comment: 10 pages, 6 figures, submitted on 31 October 2009 to Classical and
Quantum Gravity for the proceedings of 8th Edoardo Amaldi Conference on
Gravitational Wave
Calibration of the LIGO displacement actuators via laser frequency modulation
We present a frequency modulation technique for calibration of the
displacement actuators of the LIGO 4-km-long interferometric gravitational-wave
detectors. With the interferometer locked in a single-arm configuration, we
modulate the frequency of the laser light, creating an effective length
variation that we calibrate by measuring the amplitude of the frequency
modulation. By simultaneously driving the voice coil actuators that control the
length of the arm cavity, we calibrate the voice coil actuation coefficient
with an estimated 1-sigma uncertainty of less than one percent. This technique
enables a force-free, single-step actuator calibration using a displacement
fiducial that is fundamentally different from those employed in other
calibration methods.Comment: 10 pages, 5 figures, submitted to Classical and Quantum Gravit
A First Comparison Between LIGO and Virgo Inspiral Search Pipelines
This article reports on a project that is the first step the LIGO Scientific
Collaboration and the Virgo Collaboration have taken to prepare for the mutual
search for inspiral signals. The project involved comparing the analysis
pipelines of the two collaborations on data sets prepared by both sides,
containing simulated noise and injected events. The ability of the pipelines to
detect the injected events was checked, and a first comparison of how the
parameters of the events were recovered has been completed.Comment: GWDAW-9 proceeding
A first comparison of search methods for gravitational wave bursts using LIGO and Virgo simulated data
We present a comparative study of 6 search methods for gravitational wave
bursts using simulated LIGO and Virgo noise data. The data's spectra were
chosen to follow the design sensitivity of the two 4km LIGO interferometers and
the 3km Virgo interferometer. The searches were applied on replicas of the data
sets to which 8 different signals were injected. Three figures of merit were
employed in this analysis: (a) Receiver Operator Characteristic curves, (b)
necessary signal to noise ratios for the searches to achieve 50 percent and 90
percent efficiencies, and (c) variance and bias for the estimation of the
arrival time of a gravitational wave burst.Comment: GWDAW9 proceeding
A comparison of methods for gravitational wave burst searches from LIGO and Virgo
The search procedure for burst gravitational waves has been studied using 24
hours of simulated data in a network of three interferometers (Hanford 4-km,
Livingston 4-km and Virgo 3-km are the example interferometers). Several
methods to detect burst events developed in the LIGO Scientific Collaboration
(LSC) and Virgo collaboration have been studied and compared. We have performed
coincidence analysis of the triggers obtained in the different interferometers
with and without simulated signals added to the data. The benefits of having
multiple interferometers of similar sensitivity are demonstrated by comparing
the detection performance of the joint coincidence analysis with LSC and Virgo
only burst searches. Adding Virgo to the LIGO detector network can increase by
50% the detection efficiency for this search. Another advantage of a joint
LIGO-Virgo network is the ability to reconstruct the source sky position. The
reconstruction accuracy depends on the timing measurement accuracy of the
events in each interferometer, and is displayed in this paper with a fixed
source position example.Comment: LIGO-Virgo working group submitted to PR
Benefits of joint LIGO -- Virgo coincidence searches for burst and inspiral signals
We examine the benefits of performing a joint LIGO--Virgo search for
transient signals. We do this by adding burst and inspiral signals to 24 hours
of simulated detector data. We find significant advantages to performing a
joint coincidence analysis, above either a LIGO only or Virgo only search.
These include an increased detection efficiency, at a fixed false alarm rate,
to both burst and inspiral events and an ability to reconstruct the sky
location of a signal.Comment: 11 pages 8 figures, Amaldi 6 proceeding
Sensitivity Studies for Third-Generation Gravitational Wave Observatories
Advanced gravitational wave detectors, currently under construction, are
expected to directly observe gravitational wave signals of astrophysical
origin. The Einstein Telescope, a third-generation gravitational wave detector,
has been proposed in order to fully open up the emerging field of gravitational
wave astronomy. In this article we describe sensitivity models for the Einstein
Telescope and investigate potential limits imposed by fundamental noise
sources. A special focus is set on evaluating the frequency band below 10Hz
where a complex mixture of seismic, gravity gradient, suspension thermal and
radiation pressure noise dominates. We develop the most accurate sensitivity
model, referred to as ET-D, for a third-generation detector so far, including
the most relevant fundamental noise contributions.Comment: 13 pages, 7 picture
Scientific Potential of Einstein Telescope
Einstein gravitational-wave Telescope (ET) is a design study funded by the
European Commission to explore the technological challenges of and scientific
benefits from building a third generation gravitational wave detector. The
three-year study, which concluded earlier this year, has formulated the
conceptual design of an observatory that can support the implementation of new
technology for the next two to three decades. The goal of this talk is to
introduce the audience to the overall aims and objectives of the project and to
enumerate ET's potential to influence our understanding of fundamental physics,
astrophysics and cosmology.Comment: Conforms to conference proceedings, several author names correcte
A Cross-correlation method to search for gravitational wave bursts with AURIGA and Virgo
We present a method to search for transient GWs using a network of detectors
with different spectral and directional sensitivities: the interferometer Virgo
and the bar detector AURIGA. The data analysis method is based on the
measurements of the correlated energy in the network by means of a weighted
cross-correlation. To limit the computational load, this coherent analysis step
is performed around time-frequency coincident triggers selected by an excess
power event trigger generator tuned at low thresholds. The final selection of
GW candidates is performed by a combined cut on the correlated energy and on
the significance as measured by the event trigger generator. The method has
been tested on one day of data of AURIGA and Virgo during September 2005. The
outcomes are compared to the results of a stand-alone time-frequency
coincidence search. We discuss the advantages and the limits of this approach,
in view of a possible future joint search between AURIGA and one
interferometric detector.Comment: 11 pages, 6 figures, submitted to CQG special issue for Amaldi 7
Proceeding
The variable finesse locking technique
Virgo is a power recycled Michelson interferometer, with 3 km long Fabry-Perot cavities in the arms. The locking of the interferometer has been obtained with an original lock acquisition technique. The main idea is to lock the instrument away from its working point. Lock is obtained by misaligning the power recycling mirror and detuning the Michelson from the dark fringe. In this way, a good fraction of light escapes through the antisymmetric port and the power build-up inside the recycling cavity is extremely low. The benefit is that all the degrees of freedom are controlled when they are almost decoupled, and the linewidth of the recycling cavity is large. The interferometer is then adiabatically brought on to the dark fringe. This technique is referred to as variable finesse, since the recycling cavity is considered as a variable finesse Fabry-Perot. This technique has been widely tested and allows us to reach the dark fringe in few minutes, in an essentially deterministic way
- …