12 research outputs found
Development and operation of a pixel segmented liquid-filled linear array for radiotherapy quality assurance
A liquid isooctane (CH) filled ionization linear array for
radiotherapy quality assurance has been designed, built and tested. The
detector consists of 128 pixels, each of them with an area of 1.7 mm
1.7 mm and a gap of 0.5 mm. The small pixel size makes the detector ideal for
high gradient beam profiles like those present in Intensity Modulated Radiation
Therapy (IMRT) and radiosurgery. As read-out electronics we use the X-Ray Data
Acquisition System (XDAS) with the Xchip developed by the CCLRC.
Studies concerning the collection efficiency dependence on the polarization
voltage and on the dose rate have been made in order to optimize the device
operation.
In the first tests we have studied dose rate and energy dependences, and
signal reproducibility. Dose rate dependence was found lower than 2.5 % up to 5
Gy min, and energy dependence lower than 2.1 % up to 20 cm depth in
solid water. Output factors and penumbras for several rectangular fields have
been measured with the linear array and were compared with the results obtained
with a 0.125 cm air ionization chamber and radiographic film,
respectively. Finally, we have acquired profiles for an IMRT field and for a
virtual wedge. These profiles have also been compared with radiographic film
measurements. All the comparisons show a good correspondence. Signal
reproducibility was within a 2% during the test period (around three months).
The device has proved its capability to verify on-line therapy beams with
good spatial resolution and signal to noise ratio.Comment: 16 pages, 12 figures Submitted to Phys. Med. Bio
Assessment of the F@rma-ción learning resources
Resúmenes IV Congreso VetDoc de Docencia Veterinaria, León 2017 (6-7 de Julio)[ES] En el presente trabajo presentamos parte de los objetivos desarrollados en este proyecto, concretamente los relacionados con la exploración del metaconocimiento de los estudiantes en cuanto a la adquisición de competencias, tanto transversales como específicas de farmacología mediante la realización de las actividades y utilización de los recursos generados en el repositorio de F@RMA-CIÓN
eIMRT: a web platform for the verification and optimization of radiation treatment plans
The eIMRT platform is a remote distributed computing tool that provides users with Internet access to three different services: Monte Carlo optimization of treatment plans, CRT & IMRT treatment optimization, and a database of relevant radiation treatments/clinical cases. These services are accessible through a user-friendly and platform independent web page. Its flexible and scalable design focuses on providing the final users with services rather than a collection of software pieces. All input and output data (CT, contours, treatment plans and dose distributions) are handled using the DICOM format. The design, implementation, and support of the verification and optimization algorithms are hidden to the user. This allows a unified, robust handling of the software and hardware that enables these computation-intensive services. The eIMRT platform is currently hosted by the Galician Supercomputing Center (CESGA) and may be accessible upon request (there is a demo version at http://eimrt.cesga.es:8080/ eIMRT2/demo; request access in http://eimrt.cesga.es/signup.html). This paper describes all aspects of the eIMRT algorithms in depth, its user interface, and its services. Due to the flexible design of the platform, it has numerous applications including the intercenter comparison of treatment planning, the quality assurance of radiation treatments, the design and implementation of new approaches to certain types of treatments, and the sharing of information on radiation treatment techniques. In addition, the web platform and software tools developed for treatment verification and optimization have a modular design that allows the user to extend them with new algorithms. This software is not a commercial product. It is the result of the collaborative effort of different public research institutions and is planned to be distributed as an open source project. In this way, it will be available to any user; new releases will be generated with the new implemented codes or upgradesThis work was financed by Xunta de Galicia of Spain through grant PGIDT05SIN00101CT and by the European Community through the BeInGrid projectS
eIMRT: a web platform for the verification and optimization of radiation treatment plans
The eIMRT platform is a remote distributed computing tool that provides users with Internet access to three different services: Monte Carlo optimization of treatment plans, CRT & IMRT treatment optimization, and a database of relevant radiation treatments/clinical cases. These services are accessible through a user-friendly and platform independent web page. Its flexible and scalable design focuses on providing the final users with services rather than a collection of software pieces. All input and output data (CT, contours, treatment plans and dose distributions) are handled using the DICOM format. The design, implementation, and support of the verification and optimization algorithms are hidden to the user. This allows a unified, robust handling of the software and hardware that enables these computation-intensive services. The eIMRT platform is currently hosted by the Galician Supercomputing Center (CESGA) and may be accessible upon request (there is a demo version a