25 research outputs found

    Synopsis of the excitation frequencies and antenna dimensions used for electromagnetic field simulations.

    No full text
    <p>Dimensions of the bow tie antennas used for numerical EMF simulations. Magnetic field strengths ranging from 1.5 T (64 MHz) to 14.0 T (600 MHz) were applied. This approach was used to investigate specific absorption rate (SAR) distribution as a function of the excitation frequency.</p

    Synopsis of SAR simulations for frequencies ranging from 64 MHz (1.5 T) to 600 MHz (14.0 T).

    No full text
    <p>Point SAR [W/kg] distributions derived from numerical EMF simulations of an 8 channel bow tie antenna applicator using discrete MR frequencies ranging from 64 MHz (1.5 T) to 600 MHz (14.0 T). Point SAR profile along a middle line through the central axial slice of the cylindrical phantom (<b>a</b>). Point SAR distribution of the central axial slice of the cylindrical phantom (<b>b</b>). Point SAR distribution of the mid-coronal slice through the cylindrical phantom (<b>c</b>). A decrease in the size of the SAR hotspot was found for the axial and coronal view when moving to higher field strengths.</p

    Transmission fields (B<sub>1</sub><sup>+</sup>) of the hybrid applicator at 7.0 T in the human brain.

    No full text
    <p><i>In vivo</i> brain B<sub>1</sub><sup>+</sup> maps obtained from Bloch Siegert mapping of the eight independent channels of the applicator (<b>left</b>). For B<sub>1</sub><sup>+</sup> mapping an axial slice through the subject's brain was used. The colour scale is in units of 16 µT/√kW. B<sub>1</sub><sup>+</sup>map of the volunteers brain after B<sub>1</sub><sup>+</sup> shimming (<b>right</b>). The B<sub>1</sub><sup>+</sup>map shows rather uniform B<sub>1</sub><sup>+</sup>distribution.</p

    Experimental setup of the hybrid applicator used at a magnetic field strength of 7.0 T.

    No full text
    <p>Picture photograph of the eight channel TX/RX hybrid applicator implemented at 7.0T together with annotations that induce the transmission channel number (<b>left</b>). Picture photograph of the experimental setup which uses the hybrid applicator together with a cylindrical phantom at 7.0T (<b>right</b>).</p

    Experimental version of the bowtie antenna used in the hybrid applicator.

    No full text
    <p>Basic design and dimensions of the bow tie dipole building block used for MR imaging, MR thermometry and RF heating at 7.0 T (<b>a</b>). Picture photographs taken from the front, back and side of the bow tie antenna building block (<b>b</b>). Picture photograph of the cable trap design using semi rigid cable. Schematic diagram of the matching and tuning network connected to the antenna (<b>d</b>).</p

    Synopsis of the specific absorption rate distribution derived from electromagnetic field simulations.

    No full text
    <p>Specific absorption rate (SAR) hotspot diameter in the axial plane for iso-SAR 90%, iso-SAR 75%, iso-SAR 50% and iso-SAR 25% contour lines obtained from EMF simulations using discrete MR frequencies ranging from 1.5 T (64 MHz) to 14.0 T (600 MHz). (O) indicates that the whole object is included in the given iso-SAR contour. (−) indicates that no such iso-SAR value was found in the given ROI.</p

    <i>In vivo</i> imaging of the human brain and the human heart using the bow tie antennas.

    No full text
    <p>Illustration of the imaging capabilities of the hybrid TX/RX applicator driven by bow tie antennas. High spatial resolution MR images of the human brain (<b>a, b</b>). A gradient echo technique was used with a spatial resolution of: (0.5×0.5×2.0) mm<sup>3</sup>, FOV = (200×175) mm<sup>2</sup>, TR = 989 ms, TE = 25 ms, reference transmitter voltage U<sub>ref</sub> = 170 V, nominal flip angle = 35°, receiver bandwidth = 30 Hz/pixel. Minimum intensity projection derived from susceptibility weighted 3D gradient echo imaging of the human brain (<b>c</b>). Imaging parameters: spatial resolution: (0.5×0.4×1.2) mm<sup>3</sup>, FOV = (184×184) mm<sup>2</sup>, TR = 25 ms, TE = 14 ms, reference transmitter voltage U<sub>ref</sub> = 170 V, nominal flip angle = 24°, 16 slices per slab, receiver bandwidth = 120 Hz/pixel, flow compensation. Short axis view of the human heart (<b>d</b>). Images were acquired using a 2D CINE FLASH technique, FOV = (360×326) mm<sup>2</sup>, TE = 2.7 ms, TR = 5.6 ms, receiver bandwidth = 444 Hz/px, 30 cardiac phases, 8 views per segment, slice thickness 4 mm, spatial resolution: (1.4×1.4×4) mm<sup>3</sup>, nominal flip angle = 35°, reference transmitter voltage U<sub>ref</sub> = 400 V.</p

    2D steering of targeted RF heating in a phantom: simulation and experiment.

    No full text
    <p>Axial and coronal views of specific absorption rate (<b>left</b>) and temperature (<b>middle</b>) distribution derived from EMF and temperature simulations using the 8 channel applicator, a cylindrical phantom and a <sup>1</sup>H excitation frequency of 298 MHz. For comparison, a temperature map derived from MRTh acquisitions at 7T (298 MHz) using the TX/RX applicator is shown (<b>right</b>). For the experimental setup a heating period of 120 s was used. A set of phase shifts (Ch1:0°, Ch2:45°, Ch3:180°, Ch4:225°, Ch5:0°, Ch6:225°, Ch7:135°, Ch8:45°) between the bow tie antennas was used to steer the SAR and temperature hotspot towards the surface of the phantom.</p

    Targeted RF heating in a phantom: simulation and experiment.

    No full text
    <p>Axial and coronal views of specific absorption rate (<b>left</b>) and temperature (<b>middle</b>) distribution derived from EMF and temperature simulations using an 8 channel applicator together with a cylindrical phantom and a <sup>1</sup>H excitation frequency of 298 MHz. For comparison, a temperature map derived from MR thermometry of the same slice at 7T (298 MHz) using the TX/RX applicator is shown (<b>right</b>). For the experimental setup a heating period of 3 min was used. SAR and temperature hotspots were induced in the center of the phantom by using no phase shift between the bow tie antennas. P1–P4 indicate the location of the fiber optic temperature probes.</p

    RF performance of the experimental hybrid applicator.

    No full text
    <p>Noise correlation matrix obtained for the decoupling of the 8 elements included in the proposed 8 channel TX/RX applicator (<b>left</b>). Simulated B<sub>1</sub><sup>+</sup>-map in [µT/√kW] derived from a single element; channel 5 in this case (<b>middle</b>). For this purpose a transversal slice through the center of the phantom was used. For comparison the measured B<sub>1</sub><sup>+</sup>-map is shown [µT/√kW] for the same slice and bow tie antenna element (<b>right</b>).</p
    corecore