Development of a Three-Dimensional Image-Guided Needle Positioning System for Small Animal Interventions
Conventional needle positioning techniques for small animal microinjections are fraught with issues of repeatability and targeting accuracy. To improve the outcomes of these interventions a small animal needle positioning system guided by micro-computed tomography (micro-CT) imaging was developed. A phantom was developed to calibrate the geometric accuracy of micro-CT scanners to a traceable standard of measurement. Use of the phantom ensures the geometric fidelity of micro-CT images for use in image-guided interventions or other demanding quantitative applications. The design of a robot is described which features a remote center of motion architecture and is compact enough to operate within a micro-CT bore. Methods to calibrate the robot and register it to a micro-CT scanner are introduced. The performance of the robot is characterized and a mean targeting accuracy of 149 ± 41 µm estimated. The robot is finally demonstrated by completing an in vivo biomedical application
Ultra-High Field Strength MR Image-Guided Robotic Needle Delivery Device for In-Bore Small Animal Interventions
Current methods of accurate soft tissue injections in small animals are prone to many sources of error. Although efforts have been made to improve the accuracy of needle deliveries, none of the efforts have provided accurate soft tissue references. An MR image-guided robot was designed to function inside the bore of a 9.4T MR scanner to accurately deliver needles to locations within the mouse brain. The robot was designed to have no noticeable negative effects on the image quality and was localized in the MR images through the use of an MR image visible fiducial. The robot was mechanically calibrated and subsequently validated in an image-guided phantom experiment, where the mean needle targeting accuracy and needle trajectory accuracy were calculated to be 178 ± 54µm and 0.27 ± 0.65º, respectively. Finally, the device successfully demonstrated an image-guided needle targeting procedure in situ
Ultrasound-Guided Mechatronic System for Targeted Delivery of Cell-Based Cancer Vaccine Immunotherapy in Preclinical Models
Injection of dendritic cell (DC) vaccines into lymph nodes (LN) is a promising strategy for eliciting immune responses against cancer, but these injections in mouse cancer models are challenging due to the small target scale (~ 1 mm × 2 mm). Direct manual intranodal injection is difficult and can cause architectural damage to the LN, potentially disrupting crucial interactions between DC and T cells. Therefore, a second-generation ultrasound-guided mechatronic device has been developed to perform this intervention. A targeting accuracy of \u3c 500 μm will enable targeted delivery of the DCs specifically to a LN subcapsular space. The device was redesigned from its original CT-guided edition, which used a remote centre of motion architecture, to be easily integrated onto a commercially available VisualSonics imaging rail system. Subtle modifications were made to ensure simple workflow that allows for live-animal interventions that fall within the knockout periods stated in study protocols. Several calibration and registration techniques were developed in order to achieve an overall targeting accuracy appropriate for the intended application. A variety of methods to quantify the positioning accuracy of the device were investigated. The method chosen involved validating a guided injection into a tissue-mimicking phantom using ultrasound imaging post-operatively to localize the end-point position of the needle tip in the track left behind by the needle. Ultrasound-guided injections into a tissue-mimicking phantom revealed a targeting accuracy of 285 ± 94 μm for the developed robot compared to 508 ± 166 μm for a commercial-available manually-actuated injection device from VisuailSonics. The utility of the robot was also demonstrated by performing in vivo injections into the lymph nodes of mice
New Mechatronic Systems for the Diagnosis and Treatment of Cancer
Both two dimensional (2D) and three dimensional (3D) imaging modalities are useful tools for viewing the internal anatomy. Three dimensional imaging techniques are required for accurate targeting of needles. This improves the efficiency and control over the intervention as the high temporal resolution of medical images can be used to validate the location of needle and target in real time. Relying on imaging alone, however, means the intervention is still operator dependent because of the difficulty of controlling the location of the needle within the image. The objective of this thesis is to improve the accuracy and repeatability of needle-based interventions over conventional techniques: both manual and automated techniques. This includes increasing the accuracy and repeatability of these procedures in order to minimize the invasiveness of the procedure.
In this thesis, I propose that by combining the remote center of motion concept using spherical linkage components into a passive or semi-automated device, the physician will have a useful tracking and guidance system at their disposal in a package, which is less threatening than a robot to both the patient and physician. This design concept offers both the manipulative transparency of a freehand system, and tremor reduction through scaling currently offered in automated systems. In addressing each objective of this thesis, a number of novel mechanical designs incorporating an remote center of motion architecture with varying degrees of freedom have been presented. Each of these designs can be deployed in a variety of imaging modalities and clinical applications, ranging from preclinical to human interventions, with an accuracy of control in the millimeter to sub-millimeter range
Computer- and robot-assisted Medical Intervention
Medical robotics includes assistive devices used by the physician in order to
make his/her diagnostic or therapeutic practices easier and more efficient.
This chapter focuses on such systems. It introduces the general field of
Computer-Assisted Medical Interventions, its aims, its different components and
describes the place of robots in that context. The evolutions in terms of
general design and control paradigms in the development of medical robots are
presented and issues specific to that application domain are discussed. A view
of existing systems, on-going developments and future trends is given. A
case-study is detailed. Other types of robotic help in the medical environment
(such as for assisting a handicapped person, for rehabilitation of a patient or
for replacement of some damaged/suppressed limbs or organs) are out of the
scope of this chapter.Comment: Handbook of Automation, Shimon Nof (Ed.) (2009) 000-00
Engineering precision surgery: Design and implementation of surgical guidance technologies
In the quest for precision surgery, this thesis introduces several novel detection and navigation modalities for the localization of cancer-related tissues in the operating room. The engineering efforts have focused on image-guided surgery modalities that use the complementary tracer signatures of nuclear and fluorescence radiation. The first part of the thesis covers the use of “GPS-like” navigation concepts to navigate fluorescence cameras during surgery, based on SPECT images of the patient. The second part of the thesis introduces several new imaging modalities such as a hybrid 3D freehand Fluorescence and freehand SPECT imaging and navigation device. Furthermore, to improve the detection of radioactive tracer-emissions during robot-assisted laparoscopic surgery, a tethered DROP-IN gamma probe is introduced. The clinical indications that are used to evaluate the new technologies were all focused on sentinel lymph node procedures in urology (i.e. prostate and penile cancer). Nevertheless, all presented techniques are of such a nature, that they can be applied to different surgical indications, including sentinel lymph node and tumor-receptor-targeted procedures, localization the primary tumor and metastatic spread. This will hopefully contribute towards more precise, less invasive and more effective surgical procedures in the field of oncology. Crystal Photonics GmbH
Eurorad S.A.
Intuitive Surgical Inc.
KARL STORZ Endoscopie Nederland B.V.
MILabs B.V.
PI Medical Diagnostic Equipment B.V.
SurgicEye GmbH
Verb Surgical Inc.LUMC / Geneeskund
Real-time Prostate Motion Tracking For Robot-assisted Laparoscopic Radical Prostatectomy
Radical prostatectomy surgery (RP) is the gold standard for treatment of localized prostate cancer (PCa). Recently, emergence of minimally invasive techniques such as Laparoscopic Radical Prostatectomy (LRP) and Robot-Assisted Laparoscopic Radical Prostatectomy (RARP) has improved the outcomes for prostatectomy. However, it remains difficult for surgeons to make informed decisions regarding resection margins and nerve sparing since the location of the tumour within the organ is not usually visible in a laparoscopic view. While MRI enables visualization of the salient structures and cancer foci, its efficacy in LRP is reduced unless it is fused into a stereoscopic view such that homologous structures overlap. Registration of the MRI image and peri-operative ultrasound image either via visual manual alignment or using a fully automated registration can potentially be exploited to bring the pre-operative information into alignment with the patient coordinate system at the beginning of the procedure. While doing so, prostate motion needs to be compensated in real-time to synchronize the stereoscopic view with the pre-operative MRI during the prostatectomy procedure. In this thesis, two tracking methods are proposed to assess prostate rigid rotation and translation for the prostatectomy. The first method presents a 2D-to-3D point-to-line registration algorithm to measure prostate motion and translation with respect to an initial 3D TRUS image. The second method investigates a point-based stereoscopic tracking technique to compensate for rigid prostate motion so that the same motion can be applied to the pre-operative images
Navigation system based in motion tracking sensor for percutaneous renal access
Tese de Doutoramento em Engenharia BiomédicaMinimally-invasive kidney interventions are daily performed to diagnose and treat several renal
diseases. Percutaneous renal access (PRA) is an essential but challenging stage for most of these
procedures, since its outcome is directly linked to the physician’s ability to precisely visualize and
reach the anatomical target.
Nowadays, PRA is always guided with medical imaging assistance, most frequently using X-ray
based imaging (e.g. fluoroscopy). Thus, radiation on the surgical theater represents a major risk to
the medical team, where its exclusion from PRA has a direct impact diminishing the dose exposure
on both patients and physicians.
To solve the referred problems this thesis aims to develop a new hardware/software framework
to intuitively and safely guide the surgeon during PRA planning and puncturing.
In terms of surgical planning, a set of methodologies were developed to increase the certainty of
reaching a specific target inside the kidney. The most relevant abdominal structures for PRA were
automatically clustered into different 3D volumes. For that, primitive volumes were merged as a local
optimization problem using the minimum description length principle and image statistical
properties. A multi-volume Ray Cast method was then used to highlight each segmented volume.
Results show that it is possible to detect all abdominal structures surrounding the kidney, with the
ability to correctly estimate a virtual trajectory.
Concerning the percutaneous puncturing stage, either an electromagnetic or optical solution
were developed and tested in multiple in vitro, in vivo and ex vivo trials. The optical tracking solution
aids in establishing the desired puncture site and choosing the best virtual puncture trajectory.
However, this system required a line of sight to different optical markers placed at the needle base,
limiting the accuracy when tracking inside the human body. Results show that the needle tip can
deflect from its initial straight line trajectory with an error higher than 3 mm. Moreover, a complex
registration procedure and initial setup is needed.
On the other hand, a real-time electromagnetic tracking was developed. Hereto, a catheter
was inserted trans-urethrally towards the renal target. This catheter has a position and orientation
electromagnetic sensor on its tip that function as a real-time target locator. Then, a needle integrating a similar sensor is used. From the data provided by both sensors, one computes a virtual puncture
trajectory, which is displayed in a 3D visualization software. In vivo tests showed a median renal and
ureteral puncture times of 19 and 51 seconds, respectively (range 14 to 45 and 45 to 67 seconds).
Such results represent a puncture time improvement between 75% and 85% when comparing to
state of the art methods.
3D sound and vibrotactile feedback were also developed to provide additional information about
the needle orientation. By using these kind of feedback, it was verified that the surgeon tends to
follow a virtual puncture trajectory with a reduced amount of deviations from the ideal trajectory,
being able to anticipate any movement even without looking to a monitor. Best results show that 3D
sound sources were correctly identified 79.2 ± 8.1% of times with an average angulation error of
10.4º degrees. Vibration sources were accurately identified 91.1 ± 3.6% of times with an average
angulation error of 8.0º degrees.
Additionally to the EMT framework, three circular ultrasound transducers were built with a needle
working channel. One explored different manufacture fabrication setups in terms of the piezoelectric
materials, transducer construction, single vs. multi array configurations, backing and matching
material design. The A-scan signals retrieved from each transducer were filtered and processed to
automatically detect reflected echoes and to alert the surgeon when undesirable anatomical
structures are in between the puncture path. The transducers were mapped in a water tank and
tested in a study involving 45 phantoms. Results showed that the beam cross-sectional area
oscillates around the ceramics radius and it was possible to automatically detect echo signals in
phantoms with length higher than 80 mm.
Hereupon, it is expected that the introduction of the proposed system on the PRA procedure,
will allow to guide the surgeon through the optimal path towards the precise kidney target, increasing
surgeon’s confidence and reducing complications (e.g. organ perforation) during PRA. Moreover, the
developed framework has the potential to make the PRA free of radiation for both patient and surgeon
and to broad the use of PRA to less specialized surgeons.Intervenções renais minimamente invasivas são realizadas diariamente para o tratamento e
diagnóstico de várias doenças renais. O acesso renal percutâneo (ARP) é uma etapa essencial e
desafiante na maior parte destes procedimentos. O seu resultado encontra-se diretamente
relacionado com a capacidade do cirurgião visualizar e atingir com precisão o alvo anatómico.
Hoje em dia, o ARP é sempre guiado com recurso a sistemas imagiológicos, na maior parte
das vezes baseados em raios-X (p.e. a fluoroscopia). A radiação destes sistemas nas salas cirúrgicas
representa um grande risco para a equipa médica, aonde a sua remoção levará a um impacto direto
na diminuição da dose exposta aos pacientes e cirurgiões.
De modo a resolver os problemas existentes, esta tese tem como objetivo o desenvolvimento
de uma framework de hardware/software que permita, de forma intuitiva e segura, guiar o cirurgião
durante o planeamento e punção do ARP.
Em termos de planeamento, foi desenvolvido um conjunto de metodologias de modo a
aumentar a eficácia com que o alvo anatómico é alcançado. As estruturas abdominais mais
relevantes para o procedimento de ARP, foram automaticamente agrupadas em volumes 3D, através
de um problema de optimização global com base no princípio de “minimum description length” e
propriedades estatísticas da imagem. Por fim, um procedimento de Ray Cast, com múltiplas funções
de transferência, foi utilizado para enfatizar as estruturas segmentadas. Os resultados mostram que
é possível detetar todas as estruturas abdominais envolventes ao rim, com a capacidade para
estimar corretamente uma trajetória virtual.
No que diz respeito à fase de punção percutânea, foram testadas duas soluções de deteção
de movimento (ótica e eletromagnética) em múltiplos ensaios in vitro, in vivo e ex vivo. A solução
baseada em sensores óticos ajudou no cálculo do melhor ponto de punção e na definição da melhor
trajetória a seguir. Contudo, este sistema necessita de uma linha de visão com diferentes
marcadores óticos acoplados à base da agulha, limitando a precisão com que a agulha é detetada
no interior do corpo humano. Os resultados indicam que a agulha pode sofrer deflexões à medida
que vai sendo inserida, com erros superiores a 3 mm.
Por outro lado, foi desenvolvida e testada uma solução com base em sensores
eletromagnéticos. Para tal, um cateter que integra um sensor de posição e orientação na sua ponta, foi colocado por via trans-uretral junto do alvo renal. De seguida, uma agulha, integrando um sensor
semelhante, é utilizada para a punção percutânea. A partir da diferença espacial de ambos os
sensores, é possível gerar uma trajetória de punção virtual. A mediana do tempo necessário para
puncionar o rim e ureter, segundo esta trajetória, foi de 19 e 51 segundos, respetivamente
(variações de 14 a 45 e 45 a 67 segundos). Estes resultados representam uma melhoria do tempo
de punção entre 75% e 85%, quando comparados com o estado da arte dos métodos atuais.
Além do feedback visual, som 3D e feedback vibratório foram explorados de modo a fornecer
informações complementares da posição da agulha. Verificou-se que com este tipo de feedback, o
cirurgião tende a seguir uma trajetória de punção com desvios mínimos, sendo igualmente capaz
de antecipar qualquer movimento, mesmo sem olhar para o monitor. Fontes de som e vibração
podem ser corretamente detetadas em 79,2 ± 8,1% e 91,1 ± 3,6%, com erros médios de angulação
de 10.4º e 8.0 graus, respetivamente.
Adicionalmente ao sistema de navegação, foram também produzidos três transdutores de
ultrassom circulares com um canal de trabalho para a agulha. Para tal, foram exploradas diferentes
configurações de fabricação em termos de materiais piezoelétricos, transdutores multi-array ou
singulares e espessura/material de layers de suporte. Os sinais originados em cada transdutor
foram filtrados e processados de modo a detetar de forma automática os ecos refletidos, e assim,
alertar o cirurgião quando existem variações anatómicas ao longo do caminho de punção. Os
transdutores foram mapeados num tanque de água e testados em 45 phantoms. Os resultados
mostraram que o feixe de área em corte transversal oscila em torno do raio de cerâmica, e que os
ecos refletidos são detetados em phantoms com comprimentos superiores a 80 mm.
Desta forma, é expectável que a introdução deste novo sistema a nível do ARP permitirá
conduzir o cirurgião ao longo do caminho de punção ideal, aumentado a confiança do cirurgião e
reduzindo possíveis complicações (p.e. a perfuração dos órgãos). Além disso, de realçar que este
sistema apresenta o potencial de tornar o ARP livre de radiação e alarga-lo a cirurgiões menos
especializados.The present work was only possible thanks to the support by the Portuguese Science and
Technology Foundation through the PhD grant with reference SFRH/BD/74276/2010 funded by
FCT/MEC (PIDDAC) and by Fundo Europeu de Desenvolvimento Regional (FEDER), Programa
COMPETE - Programa Operacional Factores de Competitividade (POFC) do QREN
- …