Collecting system percutaneous access using real-time tracking sensors : first pig model in vivo experience
PURPOSE:
Precise needle puncture of the renal collecting system is an essential but challenging step for successful percutaneous nephrolithotomy. We evaluated the efficiency of a new real-time electromagnetic tracking system for in vivo kidney puncture.
MATERIALS AND METHODS:
Six anesthetized female pigs underwent ureterorenoscopy to place a catheter with an electromagnetic tracking sensor into the desired puncture site and ascertain puncture success. A tracked needle with a similar electromagnetic tracking sensor was subsequently navigated into the sensor in the catheter. Four punctures were performed by each of 2 surgeons in each pig, including 1 each in the kidney, middle ureter, and right and left sides. Outcome measurements were the number of attempts and the time needed to evaluate the virtual trajectory and perform percutaneous puncture.
RESULTS:
A total of 24 punctures were easily performed without complication. Surgeons required more time to evaluate the trajectory during ureteral than kidney puncture (median 15 seconds, range 14 to 18 vs 13, range 11 to 16, p=0.1). Median renal and ureteral puncture time was 19 (range 14 to 45) and 51 seconds (range 45 to 67), respectively (p=0.003). Two attempts were needed to achieve a successful ureteral puncture. The technique requires the presence of a renal stone for testing.
CONCLUSIONS:
The proposed electromagnetic tracking solution for renal collecting system puncture proved to be highly accurate, simple and quick. This method might represent a paradigm shift in percutaneous kidney access techniques.Supported by the Foundation for Science and Technology-Portugal Fellowships SFRH/BD/ 74276/2010 and SFRH/BPD/46851/2008, and Foundation for Science and Technology-Portugal Research and Development Projects PTDC/SAUBEB/ 103368/2008 and PTDC/SAU-OSM/105578/2008 FCOMP-01-0124-FEDER-011363
Validation of percutaneous puncture trajectory during renal access using 4D ultrasound reconstruction
"Progress in Biomedical Optics and Imaging, vol. 16, nr. 43"Background: An accurate percutaneous puncture is essential for disintegration and removal of renal stones. Although this procedure has proven to be safe, some organs surrounding the renal target might be accidentally perforated. This work describes a new intraoperative framework where tracked surgical tools are superimposed within 4D ultrasound imaging for security assessment of the percutaneous puncture trajectory (PPT). Methods: A PPT is first generated from the skin puncture site towards an anatomical target, using the information retrieved by electromagnetic motion tracking sensors coupled to surgical tools. Then, 2D ultrasound images acquired with a tracked probe are used to reconstruct a 4D ultrasound around the PPT under GPU processing. Volume hole-filling was performed in different processing time intervals by a tri-linear interpolation method. At spaced time intervals, the volume of the anatomical structures was segmented to ascertain if any vital structure is in between PPT and might compromise the surgical success. To enhance the volume visualization of the reconstructed structures, different render transfer functions were used. Results: Real-time US volume reconstruction and rendering with more than 25 frames/s was only possible when rendering only three orthogonal slice views. When using the whole reconstructed volume one achieved 8-15 frames/s. 3 frames/s were reached when one introduce the segmentation and detection if some structure intersected the PPT. Conclusions: The proposed framework creates a virtual and intuitive platform that can be used to identify and validate a PPT to safely and accurately perform the puncture in percutaneous nephrolithotomy.The authors acknowledge to Foundation for Science and Technology (FCT) - Portugal for the fellowships with
references: SFRH/BD/74276/2010.info:eu-repo/semantics/publishedVersio
Kidney targeting and puncturing during percutaneous nephrolithotomy: recent advances and future perspectives
Background and Purpose:
Precise needle puncture of the kidney is a challenging and essential step for successful percutaneous nephrolithotomy (PCNL). Many devices and surgical techniques have been developed to easily achieve suitable renal access. This article presents a critical review to address the methodologies and techniques for conducting kidney targeting and the puncture step during PCNL. Based on this study, research paths are also provided for PCNL procedure improvement.
METHODS:
Most relevant works concerning PCNL puncture were identified by a search of Medline/PubMed, ISI Web of Science, and Scopus databases from 2007 to December 2012. Two authors independently reviewed the studies.
RESULTS:
A total of 911 abstracts and 346 full-text articles were assessed and discussed; 52 were included in this review as a summary of the main contributions to kidney targeting and puncturing.
CONCLUSIONS:
Multiple paths and technologic advances have been proposed in the field of urology and minimally invasive surgery to improve PCNL puncture. The most relevant contributions, however, have been provided by the application of medical imaging guidance, new surgical tools, motion tracking systems, robotics, and image processing and computer graphics. Despite the multiple research paths for PCNL puncture guidance, no widely acceptable solution has yet been reached, and it remains an active and challenging research field. Future developments should focus on real-time methods, robust and accurate algorithms, and radiation free imaging techniques.The authors acknowledge Foundation for Science and Technology (FCT) for the fellowships references: SFRH/BPD/46851/2008 and SFRH/BD/74276/2010
Navigation system based in motion tracking sensor for percutaneous renal access
Tese de Doutoramento em Engenharia BiomédicaMinimally-invasive kidney interventions are daily performed to diagnose and treat several renal
diseases. Percutaneous renal access (PRA) is an essential but challenging stage for most of these
procedures, since its outcome is directly linked to the physician’s ability to precisely visualize and
reach the anatomical target.
Nowadays, PRA is always guided with medical imaging assistance, most frequently using X-ray
based imaging (e.g. fluoroscopy). Thus, radiation on the surgical theater represents a major risk to
the medical team, where its exclusion from PRA has a direct impact diminishing the dose exposure
on both patients and physicians.
To solve the referred problems this thesis aims to develop a new hardware/software framework
to intuitively and safely guide the surgeon during PRA planning and puncturing.
In terms of surgical planning, a set of methodologies were developed to increase the certainty of
reaching a specific target inside the kidney. The most relevant abdominal structures for PRA were
automatically clustered into different 3D volumes. For that, primitive volumes were merged as a local
optimization problem using the minimum description length principle and image statistical
properties. A multi-volume Ray Cast method was then used to highlight each segmented volume.
Results show that it is possible to detect all abdominal structures surrounding the kidney, with the
ability to correctly estimate a virtual trajectory.
Concerning the percutaneous puncturing stage, either an electromagnetic or optical solution
were developed and tested in multiple in vitro, in vivo and ex vivo trials. The optical tracking solution
aids in establishing the desired puncture site and choosing the best virtual puncture trajectory.
However, this system required a line of sight to different optical markers placed at the needle base,
limiting the accuracy when tracking inside the human body. Results show that the needle tip can
deflect from its initial straight line trajectory with an error higher than 3 mm. Moreover, a complex
registration procedure and initial setup is needed.
On the other hand, a real-time electromagnetic tracking was developed. Hereto, a catheter
was inserted trans-urethrally towards the renal target. This catheter has a position and orientation
electromagnetic sensor on its tip that function as a real-time target locator. Then, a needle integrating a similar sensor is used. From the data provided by both sensors, one computes a virtual puncture
trajectory, which is displayed in a 3D visualization software. In vivo tests showed a median renal and
ureteral puncture times of 19 and 51 seconds, respectively (range 14 to 45 and 45 to 67 seconds).
Such results represent a puncture time improvement between 75% and 85% when comparing to
state of the art methods.
3D sound and vibrotactile feedback were also developed to provide additional information about
the needle orientation. By using these kind of feedback, it was verified that the surgeon tends to
follow a virtual puncture trajectory with a reduced amount of deviations from the ideal trajectory,
being able to anticipate any movement even without looking to a monitor. Best results show that 3D
sound sources were correctly identified 79.2 ± 8.1% of times with an average angulation error of
10.4º degrees. Vibration sources were accurately identified 91.1 ± 3.6% of times with an average
angulation error of 8.0º degrees.
Additionally to the EMT framework, three circular ultrasound transducers were built with a needle
working channel. One explored different manufacture fabrication setups in terms of the piezoelectric
materials, transducer construction, single vs. multi array configurations, backing and matching
material design. The A-scan signals retrieved from each transducer were filtered and processed to
automatically detect reflected echoes and to alert the surgeon when undesirable anatomical
structures are in between the puncture path. The transducers were mapped in a water tank and
tested in a study involving 45 phantoms. Results showed that the beam cross-sectional area
oscillates around the ceramics radius and it was possible to automatically detect echo signals in
phantoms with length higher than 80 mm.
Hereupon, it is expected that the introduction of the proposed system on the PRA procedure,
will allow to guide the surgeon through the optimal path towards the precise kidney target, increasing
surgeon’s confidence and reducing complications (e.g. organ perforation) during PRA. Moreover, the
developed framework has the potential to make the PRA free of radiation for both patient and surgeon
and to broad the use of PRA to less specialized surgeons.Intervenções renais minimamente invasivas são realizadas diariamente para o tratamento e
diagnóstico de várias doenças renais. O acesso renal percutâneo (ARP) é uma etapa essencial e
desafiante na maior parte destes procedimentos. O seu resultado encontra-se diretamente
relacionado com a capacidade do cirurgião visualizar e atingir com precisão o alvo anatómico.
Hoje em dia, o ARP é sempre guiado com recurso a sistemas imagiológicos, na maior parte
das vezes baseados em raios-X (p.e. a fluoroscopia). A radiação destes sistemas nas salas cirúrgicas
representa um grande risco para a equipa médica, aonde a sua remoção levará a um impacto direto
na diminuição da dose exposta aos pacientes e cirurgiões.
De modo a resolver os problemas existentes, esta tese tem como objetivo o desenvolvimento
de uma framework de hardware/software que permita, de forma intuitiva e segura, guiar o cirurgião
durante o planeamento e punção do ARP.
Em termos de planeamento, foi desenvolvido um conjunto de metodologias de modo a
aumentar a eficácia com que o alvo anatómico é alcançado. As estruturas abdominais mais
relevantes para o procedimento de ARP, foram automaticamente agrupadas em volumes 3D, através
de um problema de optimização global com base no princípio de “minimum description length” e
propriedades estatísticas da imagem. Por fim, um procedimento de Ray Cast, com múltiplas funções
de transferência, foi utilizado para enfatizar as estruturas segmentadas. Os resultados mostram que
é possível detetar todas as estruturas abdominais envolventes ao rim, com a capacidade para
estimar corretamente uma trajetória virtual.
No que diz respeito à fase de punção percutânea, foram testadas duas soluções de deteção
de movimento (ótica e eletromagnética) em múltiplos ensaios in vitro, in vivo e ex vivo. A solução
baseada em sensores óticos ajudou no cálculo do melhor ponto de punção e na definição da melhor
trajetória a seguir. Contudo, este sistema necessita de uma linha de visão com diferentes
marcadores óticos acoplados à base da agulha, limitando a precisão com que a agulha é detetada
no interior do corpo humano. Os resultados indicam que a agulha pode sofrer deflexões à medida
que vai sendo inserida, com erros superiores a 3 mm.
Por outro lado, foi desenvolvida e testada uma solução com base em sensores
eletromagnéticos. Para tal, um cateter que integra um sensor de posição e orientação na sua ponta, foi colocado por via trans-uretral junto do alvo renal. De seguida, uma agulha, integrando um sensor
semelhante, é utilizada para a punção percutânea. A partir da diferença espacial de ambos os
sensores, é possível gerar uma trajetória de punção virtual. A mediana do tempo necessário para
puncionar o rim e ureter, segundo esta trajetória, foi de 19 e 51 segundos, respetivamente
(variações de 14 a 45 e 45 a 67 segundos). Estes resultados representam uma melhoria do tempo
de punção entre 75% e 85%, quando comparados com o estado da arte dos métodos atuais.
Além do feedback visual, som 3D e feedback vibratório foram explorados de modo a fornecer
informações complementares da posição da agulha. Verificou-se que com este tipo de feedback, o
cirurgião tende a seguir uma trajetória de punção com desvios mínimos, sendo igualmente capaz
de antecipar qualquer movimento, mesmo sem olhar para o monitor. Fontes de som e vibração
podem ser corretamente detetadas em 79,2 ± 8,1% e 91,1 ± 3,6%, com erros médios de angulação
de 10.4º e 8.0 graus, respetivamente.
Adicionalmente ao sistema de navegação, foram também produzidos três transdutores de
ultrassom circulares com um canal de trabalho para a agulha. Para tal, foram exploradas diferentes
configurações de fabricação em termos de materiais piezoelétricos, transdutores multi-array ou
singulares e espessura/material de layers de suporte. Os sinais originados em cada transdutor
foram filtrados e processados de modo a detetar de forma automática os ecos refletidos, e assim,
alertar o cirurgião quando existem variações anatómicas ao longo do caminho de punção. Os
transdutores foram mapeados num tanque de água e testados em 45 phantoms. Os resultados
mostraram que o feixe de área em corte transversal oscila em torno do raio de cerâmica, e que os
ecos refletidos são detetados em phantoms com comprimentos superiores a 80 mm.
Desta forma, é expectável que a introdução deste novo sistema a nível do ARP permitirá
conduzir o cirurgião ao longo do caminho de punção ideal, aumentado a confiança do cirurgião e
reduzindo possíveis complicações (p.e. a perfuração dos órgãos). Além disso, de realçar que este
sistema apresenta o potencial de tornar o ARP livre de radiação e alarga-lo a cirurgiões menos
especializados.The present work was only possible thanks to the support by the Portuguese Science and
Technology Foundation through the PhD grant with reference SFRH/BD/74276/2010 funded by
FCT/MEC (PIDDAC) and by Fundo Europeu de Desenvolvimento Regional (FEDER), Programa
COMPETE - Programa Operacional Factores de Competitividade (POFC) do QREN
Virtual reality training platform for flexible ureterorenoscopy interventions with a minimally invasive surgical robot
The total number of ureteroscopy (URS) interventions during the past years has dramatically increased due to the ongoing technological advances and the benefits associated with these techniques. However, the current URS procedure presents some drawbacks to urologic surgeons. The LITHOS project was created with the main objective of developing a surgical robotic system for flexible ureterorenoscopic lithotripsy interventions, offering a technological solution that meets the real needs of both patients and surgeons in this type of procedures. In this paper, a virtual reality environment for flexible ureterorenoscopy interventions is presented. The proposed environment provides a suitable training platform for surgeons manipulating the surgical robotic system
Development and Validation of Augmented Reality Training Simulator for Ultrasound Guided Percutaneous Renal Access
Percutaneous renal access (PCA) is a critical step in needle-based renal procedures. Traditional PCA training relies on apprenticeship, which raises concerns about patient safety and limits training opportunities. In this thesis, we reviewed simulation-based training for PCA, described the development of a novel augmented reality (AR) simulator for ultrasound (US)-guided PCA, and evaluated its validity and efficacy as a teaching tool.
Our AR simulator allows the user to practice PCA on a silicone phantom using a tracked needle and US probe emulator under the guidance of simulated US on a tablet screen. 6 Expert and 24 novice participants were recruited to evaluate the efficacy of our simulator.
Experts highly rated the realism and usefulness of our simulator, reflected by the average face validity score of 4.39 and content validity score of 4.53 on a 5-point Likert scale. Comparisons with a Mann-Whitney U test revealed significant differences (p
Our cost-effective, flexible, and easily customizable AR training simulator can provide opportunities for trainees to acquire basic skills of US-guided PCA in a safe and stress-free environment. The effectiveness of our simulator is demonstrated through strong face, content, and construct validity, indicating its value as a novel training tool
Research on real-time physics-based deformation for haptic-enabled medical simulation
This study developed a multiple effective visuo-haptic surgical engine to handle a variety of surgical manipulations in real-time. Soft tissue models are based on biomechanical experiment and continuum mechanics for greater accuracy. Such models will increase the realism of future training systems and the VR/AR/MR implementations for the operating room
Radiological Society of North America (RSNA) 3D printing Special Interest Group (SIG): Guidelines for medical 3D printing and appropriateness for clinical scenarios
Este número da revista Cadernos de Estudos Sociais estava em organização quando fomos colhidos pela morte do sociólogo Ernesto Laclau. Seu falecimento em 13 de abril de 2014 surpreendeu a todos, e particularmente ao editor Joanildo Burity, que foi seu orientando de doutorado na University of Essex, Inglaterra, e que recentemente o trouxe à Fundação Joaquim Nabuco para uma palestra, permitindo que muitos pudessem dialogar com um dos grandes intelectuais latinoamericanos contemporâneos. Assim, buscamos fazer uma homenagem ao sociólogo argentino publicando uma entrevista inédita concedida durante a sua passagem pelo Recife, em 2013, encerrando essa revista com uma sessão especial sobre a sua trajetória
A Comparison of the Thulium Fiber Laser versus Holmium:YAG Laser Lithotripsy of Upper Urinary Tract Calculi: Preliminary Results of a Randomized Prospective Clinical Trial
The Holmium:yttrium-aluminium-garnet (Ho:YAG) laser has become an indispensable tool in Endourology since its introduction in 1995, due to its safe and efficient urinary stone fragmentation capabilities. The Thulium Fiber Laser (TFL), is the latest innovation with early clinical experience suggesting improvements over the Ho:YAG. This prospective, randomized controlled study aims to compare clinical outcomes of Ho:YAG versus TFL lithotripsy for upper urinary tract calculi. Patients undergoing ureteroscopy for renal stones (size 8-20mm, density \u3e600 HU) from a single institution were prospectively recruited.
Thirty-eight patients have been recruited (Ho:YAG=20, TFL=18). The laser-on time (p=.330), total operative time (p=.849), total laser energy (p=.745), ablation speed (p=.745) and ablation efficiency (p=.745) were not statistically different between groups. Multiple trends in favor the TFL were documented, including shorter laser-on time and total operative time, improved ablation efficiency and ablation speed.
Preliminary results suggest similar outcomes for both technologies. Several parameters are trending in favor of the TFL being more efficient; however, recruitment fulfillment will be necessary to determine if there are significant differences