98 research outputs found
Transport or Store? Synthesizing Flow-based Microfluidic Biochips using Distributed Channel Storage
Flow-based microfluidic biochips have attracted much atten- tion in the EDA
community due to their miniaturized size and execution efficiency. Previous
research, however, still follows the traditional computing model with a
dedicated storage unit, which actually becomes a bottleneck of the performance
of bio- chips. In this paper, we propose the first architectural synthe- sis
framework considering distributed storage constructed tem- porarily from
transportation channels to cache fluid samples. Since distributed storage can
be accessed more efficiently than a dedicated storage unit and channels can
switch between the roles of transportation and storage easily, biochips with
this dis- tributed computing architecture can achieve a higher execution
efficiency even with fewer resources. Experimental results con- firm that the
execution efficiency of a bioassay can be improved by up to 28% while the
number of valves in the biochip can be reduced effectively.Comment: ACM/IEEE Design Automation Conference (DAC), June 201
A droplet routing technique for fault-tolerant digital microfluidic devices
AbstractâEfficient droplet routing is one of the key approaches for realizing fault-tolerant microfluidic biochips. It requires that run-time diagnosis and fault recovery can be made possible in such systems. This paper describes a droplet routing technique for a fault-tolerant digital microfluidic platform. This technique features handling of many microfluidic operations simultaneously and uses on-chip sensors for diagnosis at run-time.\ud
Once a fault is detected during the droplet routing, recovery procedures will be started-up immediately. Faulty units on the chip will be marked and isolated from the array so that the remaining droplets can still be routed along a fault-free path to their destinations. This method guarantees a non-stop fault-tolerant operation for very large microfluidic arrays.\u
Recommended from our members
BioScript: programming safe chemistry on laboratories-on-a-chip
This paper introduces BioScript, a domain-specific language (DSL) for programmable biochemistry which executes on emerging microfluidic platforms. The goal of this research is to provide a simple, intuitive, and type-safe DSL that is accessible to life science practitioners. The novel feature of the language is its syntax, which aims to optimize human readability; the technical contributions of the paper include the BioScript type system and relevant portions of its compiler. The type system ensures that certain types of errors, specific to biochemistry, do not occur, including the interaction of chemicals that may be unsafe. The compiler includes novel optimizations that place biochemical operations to execute concurrently on a spatial 2D array platform on the granularity of a control flow graph, as opposed to individual basic blocks. Results are obtained using both a cycle-accurate microfluidic simulator and a software interface to a real-world platform
Testing microelectronic biofluidic systems
According to the 2005 International Technology Roadmap for Semiconductors, the integration of emerging nondigital CMOS technologies will require radically different test methods, posing a major challenge for designers and test engineers. One such technology is microelectronic fluidic (MEF) arrays, which have rapidly gained importance in many biological, pharmaceutical, and industrial applications. The advantages of these systems, such as operation speed, use of very small amounts of liquid, on-board droplet detection, signal conditioning, and vast digital signal processing, make them very promising. However, testable design of these devices in a mass-production environment is still in its infancy, hampering their low-cost introduction to the market. This article describes analog and digital MEF design and testing method
- âŠ