Robotic catheter cardiac ablation combining ultrasound guidance and force control
Cardiac catheters allow physicians to access the inside of the heart and perform therapeutic interventions without stopping the heart or opening the chest. However, conventional manual and actuated cardiac catheters are currently unable to precisely track and manipulate the intracardiac tissue structures because of the fast tissue motion and potential for applying damaging forces. This paper addresses these challenges by proposing and implementing a robotic catheter system that uses 3D ultrasound image guidance and force control to enable constant contact with a moving target surface in order to perform interventional procedures, such as intracardiac tissue ablation. The robotic catheter system, consisting of a catheter module, ablation and force sensing end effector, drive system, and image-guidance and control system, was commanded to apply a constant force against a moving target using a position-modulated force control method. The control system uses a combination of position tracking, force feedback, and friction and backlash compensation to achieve accurate and safe catheter–tissue interactions. The catheter was able to maintain a 1 N force on a moving motion simulator target under ultrasound guidance with 0.08 N RMS error. In a simulated ablation experiment, the robotic catheter was also able to apply a consistent force on the target while maintaining ablation electrode contact with 97% less RMS contact resistance variation than a passive mechanical equivalent. In addition, the use of force control improved catheter motion tracking by approximately 20%. These results demonstrate that 3D ultrasound guidance and force tracking allow the robotic system to maintain improved contact with a moving tissue structure, thus allowing for more accurate and repeatable cardiac procedures.Engineering and Applied Science
Recommended from our members
Robotic Catheters for Beating Heart Surgery
Compliant and flexible cardiac catheters provide direct access to the inside of the heart via the vascular system without requiring clinicians to stop the heart or open the chest. However, the fast motion of the intracardiac structures makes it difficult to modify and repair the cardiac tissue in a controlled and safe manner. In addition, rigid robotic tools for beating heart surgery require the chest to be opened and the heart exposed, making the procedures highly invasive. The novel robotic catheter system presented here enables minimally invasive repair on the fast-moving structures inside the heart, like the mitral valve annulus, without the invasiveness or risks of stopped heart procedures. In this thesis, I investigate the development of 3D ultrasound-guided robotic catheters for beating heart surgery. First, the force and stiffness values of tissue structures in the left atrium are measured to develop design requirements for the system. This research shows that a catheter will experience contractile forces of 0.5 – 1.0 N and a mean tissue structure stiffness of approximately 0.1 N/mm while interacting with the mitral valve annulus. Next, this thesis presents the catheter system design, including force sensing, tissue resection, and ablation end effectors. In order to operate inside the beating heart, position and force control systems were developed to compensate for the catheter performance limitations of friction and deadzone backlash and evaluated with ex vivo and in vivo experiments. Through the addition of friction and deadzone compensation terms, the system is able to achieve position tracking with less than 1 mm RMS error and force tracking with 0.08 N RMS error under ultrasound image guidance. Finally, this thesis examines how the robotic catheter system enhances beating heart clinical procedures. Specifically, this system improves resection quality while reducing the forces experienced by the tissue by almost 80% and improves ablation performance by reducing contact resistance variations by 97% while applying a constant force on the moving tissue.Engineering and Applied Science
Modeling and Control of Steerable Ablation Catheters
Catheters are long, flexible tubes that are extensively used in vascular and cardiac interventions, e.g., cardiac ablation, coronary angiography and mitral valve annuloplasty. Catheter-based cardiac ablation is a well-accepted treatment for atrial fibrillation, a common type of cardiac arrhythmia. During this procedure, a steerable ablation catheter is guided through the vasculature to the left atrium to correct the signal pathways inside the heart and restore normal heart rhythm. The outcome of the ablation procedure depends mainly on the correct positioning of the catheter tip at the target location inside the heart and also on maintaining a consistent contact between the catheter tip and cardiac tissue. In the presence of cardiac and respiratory motions, achieving these goals during the ablation procedure is very challenging without proper 3D visualization, dexterous control of the flexible catheter and an estimate of the catheter tip/tissue contact force.
This research project provides the required basis for developing a robotics-assisted catheter manipulation system with contact force control for use in cardiac ablation procedures. The behavior of the catheter is studied in free space as well in contact with the environment to develop mathematical models of the catheter tip that are well suited for developing control systems. The validity of the proposed modeling approaches and the performance of the suggested control techniques are evaluated experimentally.
As the first step, the static force-deflection relationship for ablation catheters is described with a large-deflection beam model and an optimized pseudo-rigid-body 3R model. The proposed static model is then used in developing a control system for controlling the contact force when the catheter tip is interacting with a static environment. Our studies also showed that it is possible to estimate the tip/tissue contact force by analyzing the shape of the catheter without installing a force sensor on the catheter.
During cardiac ablation, the catheter tip is in contact with a relatively fast moving environment (cardiac tissue). Robotic manipulation of the catheter has the potential to improve the quality of contact between the catheter tip and cardiac tissue. To this end, the frequency response of the catheter is investigated and a control technique is proposed to compensate for the cardiac motion and to maintain a constant tip/tissue contact force.
Our study on developing a motion compensated robotics-assisted catheter manipulation system suggests that redesigning the actuation mechanism of current ablation catheters would provide a major improvement in using these catheters in robotics-assisted cardiac ablation procedures
Remote Navigation and Contact-Force Control of Radiofrequency Ablation Catheters
Atrial fibrillation (AF), the most common and clinically significant heart rhythm disorder, is characterized by rapid and irregular electrical activity in the upper chambers resulting in abnormal contractions. Radiofrequency (RF) cardiac catheter ablation is a minimally invasive curative treatment that aims to electrically correct signal pathways inside the atria to restore normal sinus rhythm. Successful catheter ablation requires the complete and permanent elimination of arrhythmogenic signals by delivering transmural RF ablation lesions contiguously near and around key cardiac structures. These procedures are complex and technically challenging and, even when performed by the most skilled physician, nearly half of patients undergo repeat procedures due to incomplete elimination of the arrhythmogenic pathways. This thesis aims to incorporate innovative design to improve catheter stability and maneuverability through the development of robotic platforms that enable precise placement of reproducibly durable ablation lesions.
The first part of this thesis deals with the challenges to lesion delivery imposed by cardiorespiratory motion. One of the main determinants of the delivery of durable and transmural RF lesions is the ability to define and maintain a constant contact force between the catheter tip electrode and cardiac tissue, which is hampered by the presence of cardiorespiratory motion. To address this need, I developed and evaluated a novel catheter contact-force control device. The compact electromechanical add-on tool monitors catheter-tissue contact force in real-time and simultaneously adjusts the position of a force-sensing ablation catheter within a steerable sheath to compensate for the change in contact force. In a series of in vitro and in vivo experiments, the contact-force control device demonstrated an ability to: a) maintain an average force to within 1 gram of a set level; b) reduce contact-force variation to below 5 grams (2-8-fold improvement over manual catheter intervention); c) ensure the catheter tip never lost contact with the tissue and never approached dangerous force levels; and importantly, d) deliver reproducible RF ablation lesions regardless of cardiac tissue motion, which were of the same depth and volume as lesions delivered in the absence of tissue motion.
In the second part of the thesis, I describe a novel steerable sheath and catheter robotic navigation system, which incorporates the catheter contact-force controller. The robotic platform enables precise and accurate manipulation of a remote conventional steerable sheath and permits catheter-tissue contact-force control. The robotic navigation system was evaluated in vitro using a phantom that combines stationary and moving targets within an in vitro model representing a beating heart. An electrophysiologist used the robotic system to remotely navigate the sheath and catheter tip to select targets and compared the accuracy of reaching these targets performing the same tasks manually. Robotic intervention resulted in significantly higher accuracy and significantly improved the contact-force profile between the catheter tip and moving tissue-mimicking material.
Our studies demonstrate that using available contact-force information within a robotic system can ensure precise and accurate placement of reliably transmural RF ablation lesions. These robotic systems can be valuable tools used to optimize RF lesion delivery techniques and ultimately improve clinical outcomes for AF ablation therapy
Recommended from our members
Force control of flexible catheter robots for beating heart surgery
Recent developments in cardiac catheter technology promise to allow physicians to perform most cardiac interventions without stopping the heart or opening the chest. However, current cardiac devices, including newly developed catheter robots, are unable to accurately track and interact with the fast moving cardiac tissue without applying potentially damaging forces. This paper examines the challenges of implementing force control on a flexible robotic catheter. In particular, catheter friction and backlash must be compensated when controlling tissue interaction forces. Force controller designs are introduced and evaluated experimentally in a number of configurations. The controllers are based on the inner position loop force control approach where the position trajectory is adjusted to achieve a desired force on the target. Friction and backlash compensation improved force tracking up to 86% with residual RMS errors of 0.11 N while following a prerecorded cardiac tissue trajectory with accelerations of up to 3800 mm/s. This performance provides sufficient accuracy to enable a wide range of beating heart surgical procedures.Engineering and Applied Science
Design, Development and Force Control of a Tendon-driven Steerable Catheter with a Learning-based Approach
In this research, a learning-based force control schema for tendon-driven steerable catheters with the application in robot-assisted tissue ablation procedures was proposed and validated. To this end, initially a displacement-based model for estimating the contact force between the catheter and tissue was developed. Afterward, a tendon-driven catheter was designed and developed. Next, a software-hardware-integrated robotic system for controlling and monitoring the pose of the catheter was designed and developed. Also, a force control schema was developed based on the developed contact force model as a priori knowledge. Furthermore, the position control of the tip of the catheter was performed using a learning-based inverse kinematic approach. By combining
the position control and the contact model, the force control schema was developed and validated. Validation studies were performed on phantom tissue as well as excised porcine tissue. Results of the validation studies showed that the proposed displacement-based model was 91.5% accurate in contact force prediction. Also, the system was capable of following a set of desired trajectories with an average root-mean-square error of less than 5%. Further validation studies revealed that the system could fairly generate desired static and dynamic force profiles on the phantom tissue. In summary, the proposed force control system did not necessitate the utilization of force sensors and could fairly contribute in automatizing the ablation task for robotic tissue ablation procedures
Autonomous robotic intracardiac catheter navigation using haptic vision
International audienceWhile all minimally invasive procedures involve navigating from a small incision in the skin to the site of the intervention, it has not been previously demonstrated how this can be done 10 autonomously. To show that autonomous navigation is possible, we investigated it in the hardest place to do it-inside the beating heart. We created a robotic catheter that can navigate through the blood-filled heart using wall-following algorithms inspired by positively thigmotactic animals. The catheter employs haptic vision, a hybrid sense using imaging for both touch-based surface identification and force sensing, to accomplish wall following inside the blood-filled heart. 15 Through in vivo animal experiments, we demonstrate that the performance of an autonomously-controlled robotic catheter rivals that of an experienced clinician. Autonomous navigation is a fundamental capability on which more sophisticated levels of autonomy can be built, e.g., to perform a procedure. Similar to the role of automation in fighter aircraft, such capabilities can free the clinician to focus on the most critical aspects of the procedure while providing precise and 20 repeatable tool motions independent of operator experience and fatigue
- …