Expanding the use of real-time electromagnetic tracking in radiation oncology.
In the past 10 years, techniques to improve radiotherapy delivery, such as intensity-modulated radiation therapy (IMRT), image-guided radiation therapy (IGRT) for both inter- and intrafraction tumor localization, and hypofractionated delivery techniques such as stereotactic body radiation therapy (SBRT), have evolved tremendously. This review article focuses on only one part of that evolution, electromagnetic tracking in radiation therapy. Electromagnetic tracking is still a growing technology in radiation oncology and, as such, the clinical applications are limited, the expense is high, and the reimbursement is insufficient to cover these costs. At the same time, current experience with electromagnetic tracking applied to various clinical tumor sites indicates that the potential benefits of electromagnetic tracking could be significant for patients receiving radiation therapy. Daily use of these tracking systems is minimally invasive and delivers no additional ionizing radiation to the patient, and these systems can provide explicit tumor motion data. Although there are a number of technical and fiscal issues that need to be addressed, electromagnetic tracking systems are expected to play a continued role in improving the precision of radiation delivery
Dosimetric validation of a magnetic resonance image gated radiotherapy system using a motion phantom and radiochromic film.
PurposeMagnetic resonance image (MRI) guided radiotherapy enables gating directly on the target position. We present an evaluation of an MRI-guided radiotherapy system's gating performance using an MRI-compatible respiratory motion phantom and radiochromic film. Our evaluation is geared toward validation of our institution's clinical gating protocol which involves planning to a target volume formed by expanding 5 mm about the gross tumor volume (GTV) and gating based on a 3 mm window about the GTV.MethodsThe motion phantom consisted of a target rod containing high-contrast target inserts which moved in the superior-inferior direction inside a body structure containing background contrast material. The target rod was equipped with a radiochromic film insert. Treatment plans were generated for a 3 cm diameter spherical planning target volume, and delivered to the phantom at rest and in motion with and without gating. Both sinusoidal trajectories and tumor trajectories measured during MRI-guided treatments were used. Similarity of the gated dose distribution to the planned, motion-frozen, distribution was quantified using the gamma technique.ResultsWithout gating, gamma pass rates using 4%/3 mm criteria were 22-59% depending on motion trajectory. Using our clinical standard of repeated breath holds and a gating window of 3 mm with 10% target allowed outside the gating boundary, the gamma pass rate was 97.8% with 3%/3 mm gamma criteria. Using a 3 mm window and 10% allowed excursion, all of the patient tumor motion trajectories at actual speed resulting in at least 95% gamma pass rate at 4%/3 mm.ConclusionsOur results suggest that the device can be used to compensate respiratory motion using a 3 mm gating margin and 10% allowed excursion results in conjunction with repeated breath holds. Full clinical validation requires a comprehensive evaluation of tracking performance in actual patient images, outside the scope of this study
Advances in real-time thoracic guidance systems
Substantial tissue motion: \u3e1cm) arises in the thoracic/abdominal cavity due to respiration. There are many clinical applications in which localizing tissue with high accuracy: \u3c1mm) is important. Potential applications include radiation therapy, radio frequency ablation, lung/liver biopsies, and brachytherapy seed placement. Recent efforts have made highly accurate sub-mm 3D localization of discrete points available via electromagnetic: EM) position monitoring. Technology from Calypso Medical allows for simultaneous tracking of up to three implanted wireless transponders. Additionally, Medtronic Navigation uses wired electromagnetic tracking to guide surgical tools for image guided surgery: IGS). Utilizing real-time EM position monitoring, a prototype system was developed to guide a therapeutic linear accelerator to follow a moving target: tumor) within the lung/abdomen. In a clinical setting, electromagnetic transponders would be bronchoscopically implanted into the lung of the patient in or near the tumor. These transponders would ax to the lung tissue in a stable manner and allow real-time position knowledge throughout a course of radiation therapy. During each dose of radiation, the beam is either halted when the target is outside of a given threshold, or in a later study the beam follows the target in real-time based on the EM position monitoring. We present quantitative analysis of the accuracy and efficiency of the radiation therapy tumor tracking system. EM tracking shows promise for IGS applications. Tracking the position of the instrument tip allows for minimally invasive intervention and alleviates the trauma associated with conventional surgery. Current clinical IGS implementations are limited to static targets: e.g. craniospinal, neurological, and orthopedic intervention. We present work on the development of a respiratory correlated image guided surgery: RCIGS) system. In the RCIGS system, target positions are modeled via respiratory correlated imaging: 4DCT) coupled with a breathing surrogate representative of the patient\u27s respiratory phase/amplitude. Once the target position is known with respect to the surrogate, intervention can be performed when the target is in the correct location. The RCIGS system consists of imaging techniques and custom developed software to give visual and auditory feedback to the surgeon indicating both the proper location and time for intervention. Presented here are the details of the IGS lung system along with quantitative results of the system accuracy in motion phantom, ex-vivo porcine lung, and human cadaver environments
An image-based method to synchronize cone-beam CT and optical surface tracking
open5siThe integration of in-room X-ray imaging and optical surface tracking has gained increasing importance in the field of image guided radiotherapy (IGRT). An essential step for this integration consists of temporally synchronizing the acquisition of X-ray projections and surface data. We present an image-based method for the synchronization of cone-beam computed tomography (CBCT) and optical surface systems, which does not require the use of additional hardware. The method is based on optically tracking the motion of a component of the CBCT/gantry unit, which rotates during the acquisition of the CBCT scan. A calibration procedure was implemented to relate the position of the rotating component identified by the optical system with the time elapsed since the beginning of the CBCT scan, thus obtaining the temporal correspondence between the acquisition of X-ray projections and surface data. The accuracy of the proposed synchronization method was evaluated on a motorized moving phantom, performing eight simultaneous acquisitions with an Elekta Synergy CBCT machine and the AlignRT optical device. The median time difference between the sinusoidal peaks of phantom motion signals extracted from the synchronized CBCT and AlignRT systems ranged between -3.1 and 12.9 msec, with a maximum interquartile range of 14.4 msec. The method was also applied to clinical data acquired from seven lung cancer patients, demonstrating the potential of the proposed approach in estimating the individual and daily variations in respiratory parameters and motion correlation of internal and external structures. The presented synchronization method can be particularly useful for tumor tracking applications in extracranial radiation treatments, especially in the field of patient-specific breathing models, based on the correlation between internal tumor motion and external surface surrogates.Fassi, Aurora; Schaerer, Joël; Riboldi, Marco; Sarrut, David; Baroni, GuidoFassi, Aurora; Schaerer, Joël; Riboldi, Marco; Sarrut, David; Baroni, Guid
Real-time intrafraction motion monitoring in external beam radiotherapy
© 2019 Institute of Physics and Engineering in Medicine. Radiotherapy (RT) aims to deliver a spatially conformal dose of radiation to tumours while maximizing the dose sparing to healthy tissues. However, the internal patient anatomy is constantly moving due to respiratory, cardiac, gastrointestinal and urinary activity. The long term goal of the RT community to 'see what we treat, as we treat' and to act on this information instantaneously has resulted in rapid technological innovation. Specialized treatment machines, such as robotic or gimbal-steered linear accelerators (linac) with in-room imaging suites, have been developed specifically for real-time treatment adaptation. Additional equipment, such as stereoscopic kilovoltage (kV) imaging, ultrasound transducers and electromagnetic transponders, has been developed for intrafraction motion monitoring on conventional linacs. Magnetic resonance imaging (MRI) has been integrated with cobalt treatment units and more recently with linacs. In addition to hardware innovation, software development has played a substantial role in the development of motion monitoring methods based on respiratory motion surrogates and planar kV or Megavoltage (MV) imaging that is available on standard equipped linacs. In this paper, we review and compare the different intrafraction motion monitoring methods proposed in the literature and demonstrated in real-time on clinical data as well as their possible future developments. We then discuss general considerations on validation and quality assurance for clinical implementation. Besides photon RT, particle therapy is increasingly used to treat moving targets. However, transferring motion monitoring technologies from linacs to particle beam lines presents substantial challenges. Lessons learned from the implementation of real-time intrafraction monitoring for photon RT will be used as a basis to discuss the implementation of these methods for particle RT
4D-MRI in Radiotherapy
Four-dimensional (4D) imaging provides a useful estimation of tissue motion pattern and range for radiation therapy of moving targets. 4D-CT imaging has been a standard care of practice for stereotactic body radiation therapy of moving targets. Recently, 4D-MRI has become an emerging developmental area in radiotherapy. In comparison with 4D-CT imaging, 4D-MRI provides better spatial rendering of radiotherapy targets in abdominal and pelvis regions with improved visualization of soft tissue motion. Successful implementation of 4D-MRI requires an integration of optimized acquisition protocols, advanced image reconstruction techniques, and sufficient hardware capabilities. The proposed chapter intends to introduce basic theories, current research, development, and applications of 4D-MRI in radiotherapy
Optimizing Respiratory Gated Intensity Modulated Radiation Therapy Planning and Delivery of Early-Stage Non-Small Cell Lung Cancer
Stereotactic ablative body radiotherapy (SABR) is the standard of care for inoperable early-stage non-small cell lung cancer (NSCLC) patients. However, thoracic tumours are susceptible to respiratory motion and, if unaccounted for, can potentially lead to dosimetric uncertainties. Respiratory gating is one method that limits treatment delivery to portions of the respiratory cycle, but when combined with intensity-modulated radiotherapy (IMRT), requires rigorous verification. The goal of this thesis is to optimize respiratory gated IMRT treatment planning and develop image-guided strategies to verify the dose delivery for future early-stage NSCLC patients.
Retrospective treatment plans were generated for various IMRT delivery techniques, including fixed-beam, volumetric modulated arc therapy (VMAT), and helical tomotherapy. VMAT was determined the best technique for optimizing dose conformity and efficiency.
A second treatment planning study that considered patients exhibiting significant tumour motion was conducted. Respiratory ungated and gated VMAT plans were compared. Significant decreases in V20Gy and V50%, predictors for radiation pneumonitis and irreversible fibrosis, respectively, were observed.
The predominant uncertainty of respiratory gating lies in the ability of an external surrogate marker to accurately predict internal target motion. Intrafraction triggered kV imaging was validated in a programmable motion phantom study as a method to determine how correlated the internal and external motion are during ungated and gated VMAT deliveries and to identify potential phase shifts between the motions.
KV projections acquired during gated VMAT delivery were used to reconstruct gated cone-beam CT (CBCT), providing 3D tumour position verification. Image quality and target detectability, in the presence of MV scatter from the treatment beam to the kV detector, was evaluated with various imaging parameters and under real-patient breathing motion conditions. No significant difference in image quality was observed for the CBCT acquisitions with or without the presence of MV scatter.
This thesis explores the benefits of combining respiratory gating with IMRT/VMAT for the treatment of early stage NSCLC with SABR, and evaluates advanced on-board imaging capabilities to develop dose delivery verification protocols. The results of this thesis will provide the tools necessary to confidently implement a respiratory gated radiotherapy program aimed at improving the therapeutic ratio for early-stage NSCLC
Verifying 4D gated radiotherapy using time-integrated electronic portal imaging: a phantom and clinical study
<p>Abstract</p> <p>Background</p> <p>Respiration-gated radiotherapy (RGRT) can decrease treatment toxicity by allowing for smaller treatment volumes for mobile tumors. RGRT is commonly performed using external surrogates of tumor motion. We describe the use of time-integrated electronic portal imaging (TI-EPI) to verify the position of internal structures during RGRT delivery</p> <p>Methods</p> <p>TI-EPI portals were generated by continuously collecting exit dose data (aSi500 EPID, Portal vision, Varian Medical Systems) when a respiratory motion phantom was irradiated during expiration, inspiration and free breathing phases. RGRT was delivered using the Varian RPM system, and grey value profile plots over a fixed trajectory were used to study object positions. Time-related positional information was derived by subtracting grey values from TI-EPI portals sharing the pixel matrix. TI-EPI portals were also collected in 2 patients undergoing RPM-triggered RGRT for a lung and hepatic tumor (with fiducial markers), and corresponding planning 4-dimensional CT (4DCT) scans were analyzed for motion amplitude.</p> <p>Results</p> <p>Integral grey values of phantom TI-EPI portals correlated well with mean object position in all respiratory phases. Cranio-caudal motion of internal structures ranged from 17.5–20.0 mm on planning 4DCT scans. TI-EPI of bronchial images reproduced with a mean value of 5.3 mm (1 SD 3.0 mm) located cranial to planned position. Mean hepatic fiducial markers reproduced with 3.2 mm (SD 2.2 mm) caudal to planned position. After bony alignment to exclude set-up errors, mean displacement in the two structures was 2.8 mm and 1.4 mm, respectively, and corresponding reproducibility in anatomy improved to 1.6 mm (1 SD).</p> <p>Conclusion</p> <p>TI-EPI appears to be a promising method for verifying delivery of RGRT. The RPM system was a good indirect surrogate of internal anatomy, but use of TI-EPI allowed for a direct link between anatomy and breathing patterns.</p
Optimization strategies for respiratory motion management in stereotactic body radiation therapy
Various challenges arise during the treatment of lung tumors with stereotactic body radiation therapy (SBRT), which is a form of hypofractionated high precision conformal radiation therapy delivered to small targets. The dose is applied in only a few fractions and respiratory organ and tumor motion is a source of uncertainty additional to interfractional set-up errors. Respiratory organ and tumor motion is highly patient-specific and it affects the whole radiotherapy treatment chain. In this thesis, motion management techniques for SBRT are evaluated and improved in a clinical setting. A clinical need for improvement has been present at the LMU university hospital for each issue addressed in this thesis: Initially, the usage of respiratory correlated computed tomography (4DCT), which is vital for SBRT treatment, was seen as impractical and prone to uncertainties in the data reconstruction in its current form. Therefore, the 4DCT reconstruction workflow has been improved to minimize these potential error sources. Secondly, treatment planning for tumors affected by respiratory motion was evaluated and subsequently improved. Finally, the treatment technique of respiratory gating was implemented at the clinic, which led to the need of evaluating the respiratory gating characteristics of the novel system configuration.
At first, the 4DCT reconstruction workflow used in clinical practice was investigated, as in the presence of respiratory motion the knowledge of tumor position over time is essential in SBRT treatments. Using 4DCT, the full motion range of the individual tumor can be determined. However, certain 4DCT reconstruction methods can under- or overestimate tumor motion due to limitations in the data acquisition scheme and due to the incorrect sorting of certain X-ray computed tomography (CT) image slices into different respiratory phases. As the regular clinical workflow of cycle-based sorting (CBS) without maximum inspiration detection (and therefore no clear starting point for the individual breathing cycles) seemed to be affected by these potential errors, the usage of CBS with correct maximum detection and another sorting algorithm of the respiration states, so-called local amplitude-based sorting (LAS), both have been implemented for a reduction of image artifacts and improved 4DCT quality. The three phase binning algorithms have been investigated in a phantom study (using 10 different breathing waveforms) and in a patient study (with 10 different patients). The mis-representation of the tumor volume was reduced in both implemented sorting algorithms compared to the previously used CBS approach (without correct maximum detection) in the phantom and the patient study. The clinical recommendation was the use of CBS with improved maximum detection, as too many manual interventions would be needed for the LAS workflow.
Secondly, a combination of the actual patient breathing trace during treatment, the log files generated by the linear accelerator (LINAC), and Monte Carlo (MC) four-dimensional (4D) dose calculations for each individual fraction was implemented as a 4D dose evaluation tool. This workflow was tested in a clinical environment for SBRT treatment planning on multiple CT datasets featuring: a native free-breathing 3DCT, an average intensity projection (AIP) as well as a maximum intensity projection (MIP), both obtained from the patient's 4DCT, and density overrides (DOs) in a 3DCT. This study has been carried out for 5 SBRT patients for three-dimensional conformal radiation therapy (3D-CRT) and
volumetric modulated arc therapy (VMAT) treatment plans. The dose has been recalculated on each 4DCT breathing phase according the the patient's breathing waveform and accumulated to the gross tumor volume (GTV) at the end-of-exhale (EOE) breathing phase using deformable image registration. Even though the least differences in planned and recalculated dose were found for AIP and MIP treatment planning, the results indicate a strong dependency on individual tumor motion due to the variability of breathing motion in general, and on tumor size. The combination of the patient's individual breathing trace during each SBRT fraction with 4D MC dose calculation based on the LINAC log file information leads to a good approximation of actual dose delivery.
Finally, in order to ensure precise and accurate treatment for respiratory gating techniques, the technical characteristics of the LINAC in combination with a breathing motion monitoring system as s surrogate for tumor motion have to be identified. The dose delivery accuracy and the latency of a surface imaging system in connection with a modern medical LINAC were investigated using a dynamic breathing motion phantom. The dosimetric evaluation has been carried out using a static 2D-diode array. The measurement of the dose difference between gated and ungated radiation delivery was found to be below 1% (for clinical relevant gating levels of about 30%). The beam-on latency, or time delay,
determined using radiographic films was found to be up to 851 ms±100 ms. With these known parameters, an adjustment of the pre-selected gating level or the internal target volume (ITV) margins could be made.
With the highly patient-specific character of respiratory motion, lung SBRT faces many additional challenges besides the specific issues addressed in this thesis. However, the findings of this thesis have improved clinical workflows at the Department of Radiation Oncology of the LMU University hospital. In a future perspective, a workflow using evaluation of the actual 4D dose in combination with accurate 4DCT image acquisition and specialized treatment delivery (such as respiratory gating) has the potential for a safe further reduction of treatment margins and increased sparing of organs-at-risk (OARs) in SBRT without compromising tumor dose targeting accuracy
- …