Design and Validation of a Bimanual Haptic Epidural Needle Insertion Simulator
The case experience of anesthesiologists is one of the leading causes of
accidental dural puncture and failed epidural - the most common complications
of epidural analgesia. We designed a bimanual haptic simulator to train
anesthesiologists and optimize epidural analgesia skill acquisition, and
present a validation study conducted with 15 anesthesiologists of different
competency levels from several hospitals in Israel. Our simulator emulates the
forces applied on the epidural (Touhy) needle, held by one hand, and those
applied on the Loss of Resistance (LOR) syringe, held by the second hand. The
resistance is calculated based on a model of the Epidural region layers that is
parameterized by the weight of the patient. We measured the movements of both
haptic devices, and quantified the rate of results (success, failed epidurals
and dural punctures), insertion strategies, and answers of participants to
questionnaires about their perception of the realism of the simulation. We
demonstrated good construct validity by showing that the simulator can
distinguish between real-life novices and experts. Good face and content
validity were shown in experienced users' perception of the simulator as
realistic and well-targeted. We found differences in strategies between
different level anesthesiologists, and suggest trainee-based instruction in
advanced training stages.Comment: 12 pages, 11 figure
A review of epidural simulators: Where are we today?
Thirty-one central neural blockade simulators have been implemented into clinical practice over the last thirty years either commercially or for research. This review aims to provide a detailed evaluation of why we need epidural and spinal simulators in the first instance and then draws comparisons between computer-based and manikin-based simulators. This review covers thirty-one simulators in total; sixteen of which are solely epidural simulators, nine are for epidural plus spinal or lumbar puncture simulation, and six, which are solely lumbar puncture simulators. All hardware and software components of simulators are discussed, including actuators, sensors, graphics, haptics, and virtual reality based simulators. The purpose of this comparative review is to identify the direction for future epidural simulation by outlining necessary improvements to create the ideal epidural simulator. The weaknesses of existing simulators are discussed and their strengths identified so that these can be carried forward. This review aims to provide a foundation for the future creation of advanced simulators to enhance the training of epiduralists, enabling them to comprehensively practice epidural insertion in vitro before training on patients and ultimately reducing the potential risk of harm. © 2013 IPEM
Influence of Haptic Communication on a Shared Manual Task in a Collaborative Virtual Environment
International audienceWith the advent of new haptic feedback devices, researchers are giving serious consideration to the incorporation of haptic communication in collaborative virtual environments. For instance, haptic interactions based tools can be used for medical and related education whereby students can train in minimal invasive surgery using virtual reality before approaching human subjects. To design virtual environments that support haptic communication, a deeper understanding of humans' haptic interactions is required. In this paper, human's haptic collaboration is investigated. A collaborative virtual environment was designed to support performing a shared manual task. To evaluate this system, 60 medical students participated to an experimental study. Participants were asked to perform in dyads a needle insertion task after a training period. Results show that compared to conventional training methods, a visual-haptic training improves user's collaborative performance. In addition, we found that haptic interaction influences the partners' verbal communication when sharing haptic information. This indicates that the haptic communication training changes the nature of the users' mental representations. Finally, we found that haptic interactions increased the sense of copresence in the virtual environment: haptic communication facilitates users' collaboration in a shared manual task within a shared virtual environment. Design implications for including haptic communication in virtual environments are outlined
Research on real-time physics-based deformation for haptic-enabled medical simulation
This study developed a multiple effective visuo-haptic surgical engine to handle a variety of surgical manipulations in real-time. Soft tissue models are based on biomechanical experiment and continuum mechanics for greater accuracy. Such models will increase the realism of future training systems and the VR/AR/MR implementations for the operating room
Navigation system based in motion tracking sensor for percutaneous renal access
Tese de Doutoramento em Engenharia BiomédicaMinimally-invasive kidney interventions are daily performed to diagnose and treat several renal
diseases. Percutaneous renal access (PRA) is an essential but challenging stage for most of these
procedures, since its outcome is directly linked to the physician’s ability to precisely visualize and
reach the anatomical target.
Nowadays, PRA is always guided with medical imaging assistance, most frequently using X-ray
based imaging (e.g. fluoroscopy). Thus, radiation on the surgical theater represents a major risk to
the medical team, where its exclusion from PRA has a direct impact diminishing the dose exposure
on both patients and physicians.
To solve the referred problems this thesis aims to develop a new hardware/software framework
to intuitively and safely guide the surgeon during PRA planning and puncturing.
In terms of surgical planning, a set of methodologies were developed to increase the certainty of
reaching a specific target inside the kidney. The most relevant abdominal structures for PRA were
automatically clustered into different 3D volumes. For that, primitive volumes were merged as a local
optimization problem using the minimum description length principle and image statistical
properties. A multi-volume Ray Cast method was then used to highlight each segmented volume.
Results show that it is possible to detect all abdominal structures surrounding the kidney, with the
ability to correctly estimate a virtual trajectory.
Concerning the percutaneous puncturing stage, either an electromagnetic or optical solution
were developed and tested in multiple in vitro, in vivo and ex vivo trials. The optical tracking solution
aids in establishing the desired puncture site and choosing the best virtual puncture trajectory.
However, this system required a line of sight to different optical markers placed at the needle base,
limiting the accuracy when tracking inside the human body. Results show that the needle tip can
deflect from its initial straight line trajectory with an error higher than 3 mm. Moreover, a complex
registration procedure and initial setup is needed.
On the other hand, a real-time electromagnetic tracking was developed. Hereto, a catheter
was inserted trans-urethrally towards the renal target. This catheter has a position and orientation
electromagnetic sensor on its tip that function as a real-time target locator. Then, a needle integrating a similar sensor is used. From the data provided by both sensors, one computes a virtual puncture
trajectory, which is displayed in a 3D visualization software. In vivo tests showed a median renal and
ureteral puncture times of 19 and 51 seconds, respectively (range 14 to 45 and 45 to 67 seconds).
Such results represent a puncture time improvement between 75% and 85% when comparing to
state of the art methods.
3D sound and vibrotactile feedback were also developed to provide additional information about
the needle orientation. By using these kind of feedback, it was verified that the surgeon tends to
follow a virtual puncture trajectory with a reduced amount of deviations from the ideal trajectory,
being able to anticipate any movement even without looking to a monitor. Best results show that 3D
sound sources were correctly identified 79.2 ± 8.1% of times with an average angulation error of
10.4º degrees. Vibration sources were accurately identified 91.1 ± 3.6% of times with an average
angulation error of 8.0º degrees.
Additionally to the EMT framework, three circular ultrasound transducers were built with a needle
working channel. One explored different manufacture fabrication setups in terms of the piezoelectric
materials, transducer construction, single vs. multi array configurations, backing and matching
material design. The A-scan signals retrieved from each transducer were filtered and processed to
automatically detect reflected echoes and to alert the surgeon when undesirable anatomical
structures are in between the puncture path. The transducers were mapped in a water tank and
tested in a study involving 45 phantoms. Results showed that the beam cross-sectional area
oscillates around the ceramics radius and it was possible to automatically detect echo signals in
phantoms with length higher than 80 mm.
Hereupon, it is expected that the introduction of the proposed system on the PRA procedure,
will allow to guide the surgeon through the optimal path towards the precise kidney target, increasing
surgeon’s confidence and reducing complications (e.g. organ perforation) during PRA. Moreover, the
developed framework has the potential to make the PRA free of radiation for both patient and surgeon
and to broad the use of PRA to less specialized surgeons.Intervenções renais minimamente invasivas são realizadas diariamente para o tratamento e
diagnóstico de várias doenças renais. O acesso renal percutâneo (ARP) é uma etapa essencial e
desafiante na maior parte destes procedimentos. O seu resultado encontra-se diretamente
relacionado com a capacidade do cirurgião visualizar e atingir com precisão o alvo anatómico.
Hoje em dia, o ARP é sempre guiado com recurso a sistemas imagiológicos, na maior parte
das vezes baseados em raios-X (p.e. a fluoroscopia). A radiação destes sistemas nas salas cirúrgicas
representa um grande risco para a equipa médica, aonde a sua remoção levará a um impacto direto
na diminuição da dose exposta aos pacientes e cirurgiões.
De modo a resolver os problemas existentes, esta tese tem como objetivo o desenvolvimento
de uma framework de hardware/software que permita, de forma intuitiva e segura, guiar o cirurgião
durante o planeamento e punção do ARP.
Em termos de planeamento, foi desenvolvido um conjunto de metodologias de modo a
aumentar a eficácia com que o alvo anatómico é alcançado. As estruturas abdominais mais
relevantes para o procedimento de ARP, foram automaticamente agrupadas em volumes 3D, através
de um problema de optimização global com base no princípio de “minimum description length” e
propriedades estatísticas da imagem. Por fim, um procedimento de Ray Cast, com múltiplas funções
de transferência, foi utilizado para enfatizar as estruturas segmentadas. Os resultados mostram que
é possível detetar todas as estruturas abdominais envolventes ao rim, com a capacidade para
estimar corretamente uma trajetória virtual.
No que diz respeito à fase de punção percutânea, foram testadas duas soluções de deteção
de movimento (ótica e eletromagnética) em múltiplos ensaios in vitro, in vivo e ex vivo. A solução
baseada em sensores óticos ajudou no cálculo do melhor ponto de punção e na definição da melhor
trajetória a seguir. Contudo, este sistema necessita de uma linha de visão com diferentes
marcadores óticos acoplados à base da agulha, limitando a precisão com que a agulha é detetada
no interior do corpo humano. Os resultados indicam que a agulha pode sofrer deflexões à medida
que vai sendo inserida, com erros superiores a 3 mm.
Por outro lado, foi desenvolvida e testada uma solução com base em sensores
eletromagnéticos. Para tal, um cateter que integra um sensor de posição e orientação na sua ponta, foi colocado por via trans-uretral junto do alvo renal. De seguida, uma agulha, integrando um sensor
semelhante, é utilizada para a punção percutânea. A partir da diferença espacial de ambos os
sensores, é possível gerar uma trajetória de punção virtual. A mediana do tempo necessário para
puncionar o rim e ureter, segundo esta trajetória, foi de 19 e 51 segundos, respetivamente
(variações de 14 a 45 e 45 a 67 segundos). Estes resultados representam uma melhoria do tempo
de punção entre 75% e 85%, quando comparados com o estado da arte dos métodos atuais.
Além do feedback visual, som 3D e feedback vibratório foram explorados de modo a fornecer
informações complementares da posição da agulha. Verificou-se que com este tipo de feedback, o
cirurgião tende a seguir uma trajetória de punção com desvios mínimos, sendo igualmente capaz
de antecipar qualquer movimento, mesmo sem olhar para o monitor. Fontes de som e vibração
podem ser corretamente detetadas em 79,2 ± 8,1% e 91,1 ± 3,6%, com erros médios de angulação
de 10.4º e 8.0 graus, respetivamente.
Adicionalmente ao sistema de navegação, foram também produzidos três transdutores de
ultrassom circulares com um canal de trabalho para a agulha. Para tal, foram exploradas diferentes
configurações de fabricação em termos de materiais piezoelétricos, transdutores multi-array ou
singulares e espessura/material de layers de suporte. Os sinais originados em cada transdutor
foram filtrados e processados de modo a detetar de forma automática os ecos refletidos, e assim,
alertar o cirurgião quando existem variações anatómicas ao longo do caminho de punção. Os
transdutores foram mapeados num tanque de água e testados em 45 phantoms. Os resultados
mostraram que o feixe de área em corte transversal oscila em torno do raio de cerâmica, e que os
ecos refletidos são detetados em phantoms com comprimentos superiores a 80 mm.
Desta forma, é expectável que a introdução deste novo sistema a nível do ARP permitirá
conduzir o cirurgião ao longo do caminho de punção ideal, aumentado a confiança do cirurgião e
reduzindo possíveis complicações (p.e. a perfuração dos órgãos). Além disso, de realçar que este
sistema apresenta o potencial de tornar o ARP livre de radiação e alarga-lo a cirurgiões menos
especializados.The present work was only possible thanks to the support by the Portuguese Science and
Technology Foundation through the PhD grant with reference SFRH/BD/74276/2010 funded by
FCT/MEC (PIDDAC) and by Fundo Europeu de Desenvolvimento Regional (FEDER), Programa
COMPETE - Programa Operacional Factores de Competitividade (POFC) do QREN
- …