744 research outputs found

    Left ventricular pacing vector selection by novel echo-particle imaging velocimetry analysis for optimization of quadripolar cardiac resynchronization device: A case report

    Get PDF
    Background: The availability of pacing configurations offered by quadripolar left ventricular leads could improve patients\u2019 response to cardiac resynchronization therapy; however, the selection of an optimal setting remains a challenge. Echo-particle imaging velocimetry has shown that regional anomalies of synchrony/synergy of the left ventricle are related to the alteration, reduction, or suppression of the physiological intracavitary pressure gradients. These observations are also supported by several numerical models of the left ventricle that have shown the close relationship between wall motion abnormalities, change of intraventricular flow dynamics, and abnormal distribution of forces operating on the ventricular endocardium. Case presentation: A 73-year-old white man in New York Heart Association III functional class with an ejection fraction of 27.5 % did not improve after 1 month of cardiac resynchronization therapy. Five configurations were tested and settings were defined by optimizing intraventricular flow. After 6 months, he became New York Heart Association II class with left ventricular ejection fraction of 53.2 %. Conclusions: The abnormal dynamic of pressure gradients during the cardiac cycle, through biohumoral endocrine, autocrine, and paracrine transduction, may lead to structural changes of the myocardial walls with subsequent left ventricular remodeling. The echo-particle imaging velocimetry technique may be useful for elucidating the favorable effects of cardiac resynchronization therapy on intraventricular fluid dynamics and it could be used to identify appropriate pacing setting during acute echocardiographic optimization of left pacing vector

    Impact of Estimation Uncertainty in PMU-Based Resynchronization of Continental Europe Synchronous Areas

    Get PDF
    Power system stability is a task that every system operator (SO) is required to achieve daily to ensure an uninterruptible power supply. Especially at the transmission level, for each SO it is of utmost importance to ensure proper exchange of information with other SOs, mainly in case of contingencies. However, in the last years, two major events led to the splitting of Continental Europe into two synchronous areas. These events were caused by anomalous conditions which involved in one case the fault of a transmission line and in the other a fire outage in proximity to high-voltage lines. This work analyzes these two events from the measurement point of view. In particular, we discuss the possible impact of estimation uncertainty on control decisions based on measurements of instantaneous frequency. For this purpose, we simulate five different configurations of phasor measurement units (PMUs), as characterized by different signal models, processing routines, and estimation accuracy in the presence of off-nominal or dynamic conditions. The objective is to establish the accuracy of the frequency estimates in transient conditions, more specifically during the resynchronization of the Continental Europe area. Based on this knowledge, it is possible to set more suitable conditions for resynchronization operations: the idea is to consider not only the frequency deviation between the two areas but also to take into account the respective measurement uncertainty. As confirmed by the analysis of the two real-world scenarios, such an approach would allow for minimizing the probability of adverse or even dangerous conditions such as dampened oscillations and inter-modulations

    A spatiotemporal statistical atlas of motion for the quantification of abnormal myocardial tissue velocities

    Get PDF
    International audienceIn this paper, we present a new method for the automatic comparison of myocardial motion patterns and the characterization of their degree of abnormality, based on a statistical atlas of motion built from a reference healthy population. Our main contribution is the computation of atlas-based indexes that quantify the abnormality in the motion of a given subject against a reference population, at every location in time and space. The critical computational cost inherent to the construction of an atlas is highly reduced by the definition of myocardial velocities under a small displacements hypothesis. The indexes we propose are of notable interest for the assessment of anomalies in cardiac mobility and synchronicity when applied, for instance, to candidate selection for cardiac resynchronization therapy (CRT). We built an atlas of normality using 2D ultrasound cardiac sequences from 21 healthy volunteers, to which we compared 14 CRT patients with left ventricular dyssynchrony (LVDYS). We illustrate the potential of our approach in characterizing septal flash, a specific motion pattern related to LVDYS and recently introduced as a very good predictor of response to CRT

    Implementation issues in source coding

    Get PDF
    An edge preserving image coding scheme which can be operated in both a lossy and a lossless manner was developed. The technique is an extension of the lossless encoding algorithm developed for the Mars observer spectral data. It can also be viewed as a modification of the DPCM algorithm. A packet video simulator was also developed from an existing modified packet network simulator. The coding scheme for this system is a modification of the mixture block coding (MBC) scheme described in the last report. Coding algorithms for packet video were also investigated

    Long-Term Performance of Epicardial versus Transvenous Left Ventricular Leads for Cardiac Resynchronization Therapy

    Get PDF
    Aims: to study the technical performance of epicardial left ventricular (LV) leads placed via video assisted thoracic surgery (VATS), compared to transvenously placed leads for cardiac resynchronization therapy (CRT).Methods: From 2001 until 2013, a total of 644 lead placement procedures were performed for CRT. In the case of unsuccessful transvenous LV lead placement, the patient received an epicardial LV lead. Study groups consist of 578 patients with a transvenous LV lead and 66 with an epicardial LV lead. The primary endpoint was LV-lead failure necessitating a replacement or deactivation. The secondary endpoint was energy consumption.Results: The mean follow up was 5.9 years (epicardial: 5.5 ± 3.1, transvenous: 5.9 ± 3.5). Transvenous leads failed significantly more frequently than epicardial leads with a total of 66 (11%) in the transvenous leads group vs. 2 (3%) in the epicardial lead group ( p = 0.037). Lead energy consumption was not significantly different between groups.Conclusions: Epicardial lead placement is feasible, safe and shows excellent long-term performance compared to transvenous leads. Epicardial lead placement should be considered when primary transvenous lead placement fails or as a primary lead placement strategy in challenging cases. </p

    A systematic approach to designing reliable VV optimization methodology: Assessment of internal validity of echocardiographic, electrocardiographic and haemodynamic optimization of cardiac resynchronization therapy

    Get PDF
    AbstractBackgroundIn atrial fibrillation (AF), VV optimization of biventricular pacemakers can be examined in isolation. We used this approach to evaluate internal validity of three VV optimization methods by three criteria.Methods and resultsTwenty patients (16 men, age 75±7) in AF were optimized, at two paced heart rates, by LVOT VTI (flow), non-invasive arterial pressure, and ECG (minimizing QRS duration). Each optimization method was evaluated for: singularity (unique peak of function), reproducibility of optimum, and biological plausibility of the distribution of optima.The reproducibility (standard deviation of the difference, SDD) of the optimal VV delay was 10ms for pressure, versus 8ms (p=ns) for QRS and 34ms (p<0.01) for flow.Singularity of optimum was 85% for pressure, 63% for ECG and 45% for flow (Chi2=10.9, p<0.005).The distribution of pressure optima was biologically plausible, with 80% LV pre-excited (p=0.007). The distributions of ECG (55% LV pre-excitation) and flow (45% LV pre-excitation) optima were no different to random (p=ns).The pressure-derived optimal VV delay is unaffected by the paced rate: SDD between slow and fast heart rate is 9ms, no different from the reproducibility SDD at both heart rates.ConclusionsUsing non-invasive arterial pressure, VV delay optimization by parabolic fitting is achievable with good precision, satisfying all 3 criteria of internal validity. VV optimum is unaffected by heart rate. Neither QRS minimization nor LVOT VTI satisfy all validity criteria, and therefore seem weaker candidate modalities for VV optimization. AF, unlinking interventricular from atrioventricular delay, uniquely exposes resynchronization concepts to experimental scrutiny

    Complications in recipients of cardioverter-defibrillator or cardiac resynchronization therapy: Insights from Silesian Center Defibrillator registry

    Get PDF
    Background: Current real-life information from all-comers registries from middle and east Europe about the incidence and type of complications during long-term follow-up of patients with cardioverters-defibrillators (ICD) and cardiac resynchronization devices-defibrillators (CRT-D) is still insufficient. The aim of the study was to assess the incidence and determinants of short- and long-term complications related to implantable ICD and CRT-D. Methods: We studied 1,105 recipients hospitalized in our center in 2009–2013, followed for a mean of 2.4 years (total of 2,652 patient-years). The independent association between ICD and CRT-D recipients’ and implantation-procedures’ characteristics with the incidence of complications was analyzed using multivariable Cox regression analysis. Results: In 2-month post-procedural period, 124 (11.2%) patients developed complications. Independent predictors of short-term complications (within 2 months) were: atrial fibrillation, dual chamber ICD implantation, and use of antiplatelet therapy or coumarin. Twenty-seven (2.44%) patients experienced complications, mostly lead-related (n = 21). Independent predictors of long-term complications (2–12 months after implantation) were atrial fibrillation and dual chamber ICD implantation. Conclusions: Despite significant technological progress and operators’ experience, the occurrence of complications in ICD and CRT-D recipients is still substantial. Majority of complications are recorded in the early post-implantation phase. Analysis of independent predictors of complications seem to be essential in helping to reduce adverse events in the future and strongly supports the need for routine follow-up.  

    Deep conv-attention model for diagnosing left bundle branch block from 12-lead electrocardiograms

    Full text link
    Cardiac resynchronization therapy (CRT) is a treatment that is used to compensate for irregularities in the heartbeat. Studies have shown that this treatment is more effective in heart patients with left bundle branch block (LBBB) arrhythmia. Therefore, identifying this arrhythmia is an important initial step in determining whether or not to use CRT. On the other hand, traditional methods for detecting LBBB on electrocardiograms (ECG) are often associated with errors. Thus, there is a need for an accurate method to diagnose this arrhythmia from ECG data. Machine learning, as a new field of study, has helped to increase human systems' performance. Deep learning, as a newer subfield of machine learning, has more power to analyze data and increase systems accuracy. This study presents a deep learning model for the detection of LBBB arrhythmia from 12-lead ECG data. This model consists of 1D dilated convolutional layers. Attention mechanism has also been used to identify important input data features and classify inputs more accurately. The proposed model is trained and validated on a database containing 10344 12-lead ECG samples using the 10-fold cross-validation method. The final results obtained by the model on the 12-lead ECG data are as follows. Accuracy: 98.80+-0.08%, specificity: 99.33+-0.11 %, F1 score: 73.97+-1.8%, and area under the receiver operating characteristics curve (AUC): 0.875+-0.0192. These results indicate that the proposed model in this study can effectively diagnose LBBB with good efficiency and, if used in medical centers, will greatly help diagnose this arrhythmia and early treatment
    corecore