2,974 research outputs found
Particle Physics Instrumentation
This reports summarizes the three lectures on particle physics
instrumentation given during the AEPSHEP school in November 2014 at Puri-India.
The lectures were intended to give an overview of the interaction of particles
with matter and basic particle detection principles in the context of large
detector systems like the Large Hadron Collider.Comment: 20 pages, presented at the 2nd Asia-Europe-Pacific School of
High-Energy Physics, Puri, India, 04 - 17 Nov 201
2004 ATLAS Combined Testbeam : Computation and Validation of the Electronic Calibration Constants for the Electromagnetic Calorimeter
From July to November 2004, a full slice of the ATLAS barrel detector was studied in testbeam. A complete electromagnetic barrel module was used, read by the final electronics and operated by ATLAS TDAQ software. This note describes in details the electronic calibration procedure and the cell energy reconstruction: each step of the procedure and its associated software is explicitly described. The general calibration procedure is very similar to the one applied in previous barrel and endcap standalone testbeams. Emphasis is put on tools developed in the context of the combined testbeam which can be used for commissioning and operation of the calorimeters in ATLAS. Many validation studies were performed on each calibration constant. Previously unobserved effects such as the FEB temperature dependence of some constants were observed. Overall, the calibration performances are at the expected level
Energy Linearity and Resolution of the ATLAS Electromagnetic Barrel Calorimeter in an Electron Test-Beam
A module of the ATLAS electromagnetic barrel liquid argon calorimeter was
exposed to the CERN electron test-beam at the H8 beam line upgraded for
precision momentum measurement. The available energies of the electron beam
ranged from 10 to 245 GeV. The electron beam impinged at one point
corresponding to a pseudo-rapidity of eta=0.687 and an azimuthal angle of
phi=0.28 in the ATLAS coordinate system. A detailed study of several effects
biasing the electron energy measurement allowed an energy reconstruction
procedure to be developed that ensures a good linearity and a good resolution.
Use is made of detailed Monte Carlo simulations based on Geant which describe
the longitudinal and transverse shower profiles as well as the energy
distributions. For electron energies between 15 GeV and 180 GeV the deviation
of the measured incident electron energy over the beam energy is within 0.1%.
The systematic uncertainty of the measurement is about 0.1% at low energies and
negligible at high energies. The energy resolution is found to be about 10%
sqrt(E) for the sampling term and about 0.2% for the local constant term
Measurement of the branching ratios of the decays Xi0 --> Sigma+ e- nubar and anti-Xi0 --> anti-Sigma+ e+ nu
From 56 days of data taking in 2002, the NA48/1 experiment observed 6316 Xi0
--> Sigma+ e- nubar candidates (with the subsequent Sigma+ --> p pi0 decay) and
555 anti-Xi0 --> anti-Sigma+ e+ nu candidates with background contamination of
215+-44 and 136+-8 events, respectively. From these samples, the branching
ratios BR(Xi0 --> Sigma+ e- nubar)= (2.51+-0.03stat+-0.09syst)E(-4) and
BR(anti-Xi0 --> anti-Sigma+ e+ nu)= (2.55+-0.14stat+-0.10syst)E(-4) were
measured allowing the determination of the CKM matrix element |Vus| =
0.209+0.023-0.028. Using the Particle Data Group average for |Vus| obtained in
semileptonic kaon decays, we measured the ratio g1/f1 = 1.20+-0.05 of the
axial-vector to vector form factors.Comment: 16 pages, 11 figures Submitted to Phys.Lett.
Measurement of K^0_e3 form factors
The semileptonic decay of the neutral K meson, KL -> pi e nu (Ke3), was used
to study the strangeness-changing weak interaction of hadrons. A sample of 5.6
million reconstructed events recorded by the NA48 experiment was used to
measure the Dalitz plot density. Admitting all possible Lorentz-covariant
couplings, the form factors for vector (f_+(q^2)), scalar (f_S) and tensor
(f_T) interactions were measured. The linear slope of the vector form factor
lambda_+ = 0.0284+-0.0007+-0.0013 and values for the ratios |f_S/f_+(0)| =
0.015^{+0.007}_{-0.010}+-0.012 and |f_T/f_+(0)| = 0.05^{+0.03}_{-0.04}+-0.03
were obtained. The values for f_S and f_T are consistent with zero. Assuming
only Vector-Axial vector couplings, lambda_+ = 0.0288+-0.0004+-0.0011 and a
good fit consistent with pure V-A couplings were obtained. Alternatively, a fit
to a dipole form factor yields a pole mass of M = 859+-18 MeV, consistent with
the K^*(892) mass.Comment: 16 pages, 7 figures. submitted to Phys. Lett.
Position resolution and particle identification with the ATLAS EM calorimeter
In the years between 2000 and 2002 several pre-series and series modules of
the ATLAS EM barrel and end-cap calorimeter were exposed to electron, photon
and pion beams. The performance of the calorimeter with respect to its finely
segmented first sampling has been studied. The polar angle resolution has been
found to be in the range 50-60 mrad/sqrt(E (GeV)). The neutral pion rejection
has been measured to be about 3.5 for 90% photon selection efficiency at pT=50
GeV/c. Electron-pion separation studies have indicated that a pion fake rate of
(0.07-0.5)% can be achieved while maintaining 90% electron identification
efficiency for energies up to 40 GeV.Comment: 32 pages, 22 figures, to be published in NIM
- …