2 research outputs found

    Automated speckle tracking algorithm to aid on-axis imaging in echocardiography

    No full text
    Obtaining a “correct” view in echocardiography is a subjective process in which an operator attempts to obtain images conforming to consensus standard views. Real-time objective quantification of image alignment may assist less experienced operators, but no reliable index yet exists. We present a fully automated algorithm for detecting incorrect medial/lateral translation of an ultrasound probe by image analysis. The ability of the algorithm to distinguish optimal from sub-optimal four-chamber images was compared to that of specialists—the current “gold-standard.” The orientation assessments produced by the automated algorithm correlated well with consensus visual assessments of the specialists (r=0.87r=0.87) and compared favourably with the correlation between individual specialists and the consensus, 0.82±0.09. Each individual specialist’s assessments were within the consensus of other specialists, 75±14% of the time, and the algorithm’s assessments were within the consensus of specialists 85% of the time. The mean discrepancy in probe translation values between individual specialists and their consensus was 0.97±0.87??cm, and between the automated algorithm and specialists’ consensus was 0.92±0.70??cm. This technology could be incorporated into hardware to provide real-time guidance for image optimisation—a potentially valuable tool both for training and quality control.</p