Article thumbnail

Calibration of the TonoLab Tonometer in Mice with Spontaneous or Experimental Glaucoma

By Mary E. Pease, Frances E. Cone, Scott Gelman, Janice L. Son and Harry A. Quigley

Abstract

Measurements of intraocular pressure were made in normal and glaucomatous mice with the TonoLab tonometer, comparing its reading when pressures were set manometrically from 10 to 50 mm Hg. The TonoLab accurately reflected the set pressure, with a slight overestimation of 1 to 3 mm Hg

Topics: Articles
Publisher: Association for Research in Vision and Ophthalmology, Inc.
OAI identifier: oai:pubmedcentral.nih.gov:3053110
Provided by: PubMed Central
Download PDF:
Sorry, we are unable to provide the full text but you may find it at the following location(s):
  • http://www.pubmedcentral.nih.g... (external link)

  • To submit an update or takedown request for this paper, please submit an Update/Correction/Removal Request.