Effects of Beam Startup Characteristics on Dose Delivery Accuracy at Low Monitor Units in Step-and-Shoot Intensity Modulated Radiation Therapy

By Kazunori Fujimoto, Kunihiko Tateoka, Yuji Yaegashi, Katsumi Shima, Junji Suzuki, Yuichi Saito, Akihiro Nakata, Takuya Nakazawa, Tadanori Abe and Masaki Yano

Abstract

Intensity modulated radiation therapy (IMRT) is a highly accurate technique that is usually implemented in either dy-namic or step-and-shoot fashion with many segments each having low monitor units (MUs). The present study evalu-ated the effects of beam startup characteristics on the dose delivery accuracy for each segment at low MUs for step-and-shoot IMRT with an Elekta Precise accelerator at the highest dose rates. We used a two-dimensional semi-conductor detector for the dose measurements. The field size of each segment was assumed to be 20 × 20 cm2 and each segment was set to deliver 1- 10 MUs. Our results show a variation in dose delivery accuracy between segments for the same IMRT beam, which can be attributed to the beam startup characteristics. This variability is attributed to the changes in the transient changes in the temperatures of the electron gun filament and the magnetron. That is, the tran-sient increase in the temperature of the filament leads to increasing doses with time and that of the magnetron leads to decreasing doses with time during the first few MUs

Topics: Startup Performance, Step-and-Shoot IMRT, Low Monitor Units
Year: 2012
OAI identifier: oai:CiteSeerX.psu:10.1.1.1024.6518
Provided by: CiteSeerX
Download PDF:
Sorry, we are unable to provide the full text but you may find it at the following location(s):
  • http://file.scirp.org/pdf/IJMP... (external link)
  • http://file.scirp.org/pdf/IJMP... (external link)
  • http://citeseerx.ist.psu.edu/v... (external link)
  • Suggested articles


    To submit an update or takedown request for this paper, please submit an Update/Correction/Removal Request.