Location of Repository

Dosimetric verification of the anisotropic analytical algorithm for radiotherapy treatment planning

By C. Bragg and J. Conway

Abstract

BACKGROUND AND PURPOSE: \ud \ud To investigate the accuracy of photon dose calculations performed by the Anisotropic Analytical Algorithm, in homogeneous and inhomogeneous media and in simulated\ud treatment plans.\ud \ud MATERIALS AND METHODS: \ud \ud Predicted dose distributions were compared with ionisation chamber and film\ud measurements for a series of increasingly complex situations. Initially, simple and complex fields in a\ud homogeneous medium were studied. The effect of inhomogeneities was investigated using a range of\ud phantoms constructed of water, bone and lung substitute materials. Simulated treatment plans were\ud then produced using a semi-anthropomorphic phantom and the delivered doses compared to the doses predicted by the Anisotropic Analytical Algorithm.\ud \ud \ud RESULTS: \ud \ud In a homogeneous medium, agreement was found to be within 2% dose or 2mm dta in most\ud instances. In the presence of heterogeneities, agreement was generally to within 2.5%. The simulated\ud treatment plan measurements agreed to within 2.5% or 2mm.\ud Conclusions: The accuracy of the algorithm was found to be satisfactory at 6MV and 10MV both in homogeneous and inhomogeneous situations and in the simulated treatment plans. The algorithm was more accurate than the Pencil Beam Convolution model, particularly in the presence of low density heterogeneities.\u

Publisher: Elsevier Ireland Ltd.
Year: 2006
OAI identifier: oai:eprints.whiterose.ac.uk:3703

Suggested articles

Preview


To submit an update or takedown request for this paper, please submit an Update/Correction/Removal Request.