Design and evaluation of an RF front-end for 9.4 T human MRI

Abstract

At the field strength of 9.4 T, the highest field currently available for human MRI, the wavelength of the MR signals is significantly shorter than the size of the examined structures. Even more than at 7 T, constructive and destructive interferences cause strong inhomogeneities of the B1 field produced by a volume coil, causing shading over large parts of the image. Specialized radio frequency hardware and B1 management methods are required to obtain high-quality images that take full advantage of the high field strength. Here, the design and characteristics of a radio frequency front-end especially developed for proton imaging at 9.4 T are presented. In addition to a 16-channel transceiver array coil, capable of volume transmit mode and independent signal reception, it consists of custom built low noise preamplifiers and TR switches. Destructive interference patterns were eliminated, in virtually the entire brain, using a simple in situ radio frequency phase shimming technique. After mapping the Bmath image profile of each transmit channel, a numerical algorithm was used to calculate the appropriate transmit phase offsets needed to obtain a homogeneous excitation field over a user defined region. Between two and three phase settings are necessary to obtain homogeneous images over the entire brain

    Similar works

    Full text

    thumbnail-image

    Enlighten

    redirect
    Last time updated on 12/10/2016

    This paper was published in Enlighten.

    Having an issue?

    Is data on this page outdated, violates copyrights or anything else? Report the problem now and we will take corresponding actions after reviewing your request.